选择离子流管质谱在实时大气监测中的应用。

Application of selected ion flow tube mass spectrometry to real-time atmospheric monitoring.

机构信息

Syft Technologies Ltd., 3 Craft Place, Christchurch, New Zealand.

出版信息

Rapid Commun Mass Spectrom. 2010 Jun 30;24(12):1763-9. doi: 10.1002/rcm.4574.

Abstract

Data are presented for real-time atmospheric monitoring of volatile organic chemicals (VOCs) in air using selected ion flow tube mass spectrometry (SIFT-MS) technology. These measurements were made by one of the new generation of SIFT-MS instruments. Results are shown for five VOCs that were continually monitored from a stationary sampling point over a 4-day period: ethene, ethanol, 1,3-butadiene, benzene and toluene. All analytes except ethene in the study have at least two simultaneous and independent measures of concentration. These results demonstrate the great advances in SIFT-MS that have been made in recent years. 1,3-Butadiene is measured at a concentration of 9 pptv with a precision of 44%. For a 1-s integration time, a detection limit of 50 pptv is achieved. Instrument sensitivities are reported for all five analytes.