利用眼前节光学相干断层扫描评估波士顿 I 型角膜假体-供体角膜界面的稳定性。

Evaluation of the stability of Boston type I keratoprosthesis-donor cornea interface using anterior segment optical coherence tomography.

机构信息

The New York Eye and Ear Infirmary, New York, USA.

出版信息

Cornea. 2010 Sep;29(9):1031-5. doi: 10.1097/ICO.0b013e3181ca2ea5.

Abstract

PURPOSE

To evaluate the anatomic stability of an implanted Boston type I keratoprosthesis (KPro)-donor cornea interface and assess the presence or absence of a potential space (gap) between the KPro front plate and donor cornea using anterior segment optical coherence tomography (AS-OCT). The presence of a gap would raise concerns of a possible pathway for the exchange of extraocular fluid with the anterior chamber.

METHODS

Fifteen eyes implanted with a Boston type I KPro were studied by the noncontact technique of AS-OCT (AC Cornea OCT prototype; OTI, Canada). All the KPro devices had been implanted at least 4 weeks before the study (mean: 7 months, range: 1-22 months). Eight eyes had aphakic Kpros, and the other 7 had pseudophakic implants. Anesthetized eyes were imaged before and during pressure application using sterile cotton-tip applicators. Pressure was applied for 10 seconds on the nasal or temporal side of the eye. Images were analyzed for any possible changes in the KPro-donor cornea interface during the application of pressure.

RESULTS

Of 15 eyes, 10 had the threaded front plate model with a T-shaped silhouette and corrugated sides, whereas 5 had the threadless type with a T-shaped silhouette and smooth sides on cross-sectional optical coherence tomography. Of the 15 eyes, 2 revealed a gap between the front plate and the surface of the donor cornea. The rest revealed no gaps. With pressure, none of the eyes, including the 2 with gaps, demonstrated any change in the KPro-donor cornea interface during dynamic imaging (eg, gaping or evidence of fluid escape along the KPro-donor cornea borders). In all eyes, the position of the titanium locking ring was visible and verified to be in an adequate position.

CONCLUSIONS

The implanted KPro-donor cornea interface seems to be stable dynamically using AS-OCT. A gap that has been documented with this imaging tool showed neither gaping nor escape of anterior chamber fluid during dynamic cross-sectional imaging. Further studies will be needed to assess the cause of these gaps and to determine if they are possibly related to the hydration status of the cornea, suture tension on the donor cornea, or other factors that remain to be determined.

摘要

目的

使用眼前节光学相干断层扫描(AS-OCT)评估植入的波士顿 I 型角膜假体(KPro)-供体角膜界面的解剖稳定性,并评估 KPro 前板与供体角膜之间是否存在潜在空间(间隙)。间隙的存在会引起对眼外液与前房交换的可能途径的关注。

方法

使用非接触式眼前节 OCT(AC Cornea OCT 原型;加拿大 OTI)对 15 只植入波士顿 I 型 KPro 的眼睛进行研究。所有 KPro 装置均在研究前至少植入 4 周(平均:7 个月,范围:1-22 个月)。8 只眼为无晶状体 Kpro,其余 7 只为假晶状体植入。在使用无菌棉签施加压力前后对麻醉眼进行成像。在眼睛的鼻侧或颞侧施加压力 10 秒。分析图像以了解在施加压力时 KPro-供体角膜界面是否发生任何变化。

结果

在 15 只眼中,有 10 只具有带 T 形轮廓和波纹状侧面的螺纹前板模型,而 5 只具有无螺纹类型,其具有 T 形轮廓和横截面光学相干断层扫描的光滑侧面。在 15 只眼中,有 2 只显示出前板和供体角膜表面之间存在间隙。其余没有间隙。在施加压力时,包括有间隙的 2 只眼睛在内,所有眼睛在动态成像过程中均未显示 KPro-供体角膜界面发生任何变化(例如,张开或沿 KPro-供体角膜边界有液体逸出的证据)。在所有眼睛中,都可以看到钛锁定环的位置,并确认其处于适当位置。

结论

使用 AS-OCT,植入的 KPro-供体角膜界面在动态方面似乎是稳定的。使用这种成像工具记录的间隙在动态横截面成像过程中既没有张开,也没有前房液体逸出。需要进一步的研究来评估这些间隙的原因,并确定它们是否可能与角膜的水合状态、供体角膜上的缝线张力或其他有待确定的因素有关。