The in vivo relationship between cross-sectional area and CT dose index in abdominal multidetector CT with automatic exposure control.

作者信息

Meeson S, Alvey C M, Golding S J

机构信息

Radiology Group, Nuffield Department of Surgery, University of Oxford, John Radcliffe Hospital, Oxford OX3 9DU, UK.

出版信息

J Radiol Prot. 2010 Jun;30(2):139-47. doi: 10.1088/0952-4746/30/2/003. Epub 2010 Jun 9.

Abstract

The relationship between patient cross-sectional area and both volume CT dose index (CTDI) and dose length product was explored for abdominal CT in vivo, using a 16 multidetector row CT (MDCT) scanner with automatic exposure control. During a year-long retrospective survey of patients with MDCT for symptoms of abdominal sepsis, cross-sectional areas were estimated using customised ellipses at the level of the middle of vertebra L3. The relationship between cross-sectional area and the exposure parameters was explored. Scans were performed using a LightSpeed 16 (GE Healthcare Medical Systems, Milwaukee, WI) operated with tube current modulation. From a survey of 94 patients it was found that the CTDI increased with the increase in patient cross-sectional area. The relationship was logarithmic rather than linear, with a least-squares fit to the data (R(2) = 0.80). For abdominal CT the cross-sectional area gave a measure of patient size based on the region of the body to be exposed. Exposure parameters increased with increasing cross-sectional area and the greater radiation exposure of larger patients was partly a consequence of their size. Given increasing obesity levels we believe that cross-sectional area and scan length should be added to future dose surveys, allowing patient size to be considered as a factor of relevance when examining population doses.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验