Kinnstaetter K, Lohmann A W, Schwider J, Streibl N
Appl Opt. 1988 Dec 15;27(24):5082-9. doi: 10.1364/AO.27.005082.
The accuracy of phase shifting interferometers is impaired by mechanical drifts and vibrations, intensity variations, nonlinearities of the photoelectric detection device, and, most seriously, by inaccuracies of the reference phase shifter. The phase shifting procedure enables the detection of most of the errors listed above by a special Lissajous display technique described here. Furthermore, it is possible to correct phase shifter inaccuracies by using an iterative process relying solely on the interference pattern itself and the Fourier sums used in phase shifting interferometry.
相移干涉仪的精度会受到机械漂移和振动、强度变化、光电检测装置的非线性的影响,而最严重的是受到参考移相器不准确的影响。通过此处描述的一种特殊的李萨如图形显示技术,相移过程能够检测上述大多数误差。此外,仅依靠干涉图样本身以及相移干涉测量中使用的傅里叶和,通过迭代过程就有可能校正移相器的不准确之处。