We compared the inter-day reproducibility of post-occlusive reactive hyperemia (PORH) assessed by single-point laser Doppler flowmetry (LDF) and laser speckle contrast analysis (LSCI), and the reproducibility of local thermal hyperemia (LTH) assessed by LDF, laser Doppler imaging (LDI) and LSCI. We also tested whether skin blood flow assessment by LDF and by LSCI are correlated.
Skin blood flow was evaluated during PORH and LTH using LDF, LDI (for LTH only) and LSCI on the forearms of healthy volunteers, at a 7day interval. Data are expressed as cutaneous vascular conductance (CVC), as a function of baseline and scaled to the thermal plateau. Reproducibility is expressed as within subject coefficients of variation (CV, in %) and intra-class correlation coefficients (ICC).
Twenty-eight healthy participants were enrolled in this study. The reproducibility of the PORH peak CVC was better when assessed with LSCI compared to LDF (CV=8%; ICC=0.76 and CV=30%; ICC=0.54, respectively). Inter-day reproducibility of the LTH plateau was better when assessed with LSCI or LDI than LDF (CV=15%, ICC=0.66; CV=17%, ICC=0.51 and CV=42%, ICC=0.28 respectively). Finally, we observed significant correlation between simultaneous LDF and LSCI measurements of the PORH peak CVC (R=0.54; p=0.001).
The recently developed LSCI technique showed very good inter-day reproducibility for assessing PORH and LTH. Moreover, we showed significant correlation between LSCI and single-point LDF for PORH. However, more data are needed to evaluate the linearity between the LSCI signal and skin blood flow.
我们比较了单点激光多普勒流量仪(LDF)和激光散斑对比分析(LSCI)评估的阻断后充血反应(PORH)以及 LDF、激光多普勒成像(LDI)和 LSCI 评估的局部热充血(LTH)的日内重复性。我们还测试了 LDF 和 LSCI 评估的皮肤血流是否相关。
在 7 天的间隔内,使用 LDF、LDI(仅用于 LTH)和 LSCI 在前臂评估健康志愿者的 PORH 和 LTH 期间的皮肤血流。数据表示为皮肤血管传导率(CVC),作为基线的函数,并按热平台进行缩放。重复性表示为个体内变异系数(CV,以%表示)和组内相关系数(ICC)。
本研究共纳入 28 名健康参与者。与 LDF 相比,LSCI 评估的 PORH 峰值 CVC 重复性更好(CV=8%;ICC=0.76 和 CV=30%;ICC=0.54)。LSCI 或 LDI 评估的 LTH 平台的日内重复性优于 LDF(CV=15%,ICC=0.66;CV=17%,ICC=0.51 和 CV=42%,ICC=0.28)。最后,我们观察到 PORH 峰值 CVC 的同时 LDF 和 LSCI 测量之间存在显著相关性(R=0.54;p=0.001)。
最近开发的 LSCI 技术在评估 PORH 和 LTH 方面表现出非常好的日内重复性。此外,我们还表明 LSCI 和单点 LDF 之间存在 PORH 的显著相关性。然而,需要更多的数据来评估 LSCI 信号与皮肤血流之间的线性关系。