Suppr超能文献

[弯曲对喷丸处理和抛光的接骨板的影响]

[Effect of bending on shot peened and polished osteosynthesis plates].

作者信息

Starker M, Fröhling M, Hirsch T

机构信息

Orthopädische Universitätsklinik, Friedrichsheim, Frankfurt.

出版信息

Biomed Tech (Berl). 1991 Mar;36(3):56-9. doi: 10.1515/bmte.1991.36.3.56.

Abstract

Shot peening can increase the fatigue strength of commercially available surgical plates made of 1.4435 alloy by 40% even in a corrosive environment. Our investigations show that residual stresses resulting from shot peening are reduced by additional bending of the plates. In such plates smaller tensile residual stresses were found than after polishing of the plates. Bending of polished plates results in considerable tensile residual stresses. The hardening achieved by shot peening is not reduced by bending. As the fatigue strength of soft materials depends mainly on the hardening and less on the residual stresses, only little influence of the changed residual stresses on the fatigue strength can be expected. Shot peening of surgical implants thus means an improvement in quality.

摘要

喷丸处理可以使市售的由1.4435合金制成的外科手术钢板的疲劳强度即使在腐蚀性环境中也提高40%。我们的研究表明,喷丸处理产生的残余应力会因钢板的额外弯曲而降低。在这种钢板中,发现的拉伸残余应力比钢板抛光后更小。抛光钢板的弯曲会导致相当大的拉伸残余应力。喷丸处理所实现的硬化不会因弯曲而降低。由于软材料的疲劳强度主要取决于硬化,而较少取决于残余应力,因此可以预期残余应力的变化对疲劳强度的影响很小。因此,外科植入物的喷丸处理意味着质量的提高。