Measurement of systolic blood pressure using pulse oximetry during helicopter flight.

作者信息

Talke P O

机构信息

Department of Anesthesiology, University of Texas Medical Branch, Shriners Burns Institute, Galveston 77550.

出版信息

Crit Care Med. 1991 Jul;19(7):934-7. doi: 10.1097/00003246-199107000-00018.

Abstract

OBJECTIVE

Monitoring of vital signs in critically ill patients during helicopter flight is difficult because of the noise and vibrations of the aircraft. We evaluated the use of a pulse oximeter to measure systolic BP intraflight.

DESIGN

Systolic BP measured by pulse oximetry was compared with systolic BP measured by the direct intra-arterial and the arterial occlusion methods intraflight. Systolic BP by pulse oximetry was measured by observing the return of the plethysmographic waveform of the pulse oximeter as the BP cuff ipsilateral to the pulse oximeter probe was slowly deflated. Arterial occlusion pressure was measured by observing the return of the intraarterial waveform as the BP cuff ipsilateral to the arterial cannula was slowly deflated.

SETTING

The study was performed during patient transport, intraflight.

PATIENTS

Ten critically ill patients were studied.

INTERVENTIONS

None

MEASUREMENTS AND MAIN RESULTS

Seventy-three sets of measurements were recorded. The best correlation (r2 = .99) was found between pulse oximetry and the arterial occlusion method. The indirect methods correlated better with each other than with direct intraarterial measurements. The noise and the vibrations of the helicopter did not significantly interfere with the operation of the pulse oximeter.

CONCLUSIONS

We conclude that a pulse oximeter that displays a plethysmographic waveform can accurately measure systolic BP intraflight.

摘要

目的

由于直升机飞行时的噪音和振动,对危重症患者进行生命体征监测存在困难。我们评估了使用脉搏血氧仪在飞行中测量收缩压的情况。

设计

将飞行中通过脉搏血氧测定法测得的收缩压与通过直接动脉内测量法和动脉闭塞法测得的收缩压进行比较。通过观察与脉搏血氧仪探头同侧的血压袖带缓慢放气时脉搏血氧仪容积描记波形的恢复情况来测量脉搏血氧测定法测得的收缩压。通过观察与动脉插管同侧的血压袖带缓慢放气时动脉内波形的恢复情况来测量动脉闭塞压。

地点

该研究在患者飞行转运过程中进行。

患者

研究了10例危重症患者。

干预措施

测量与主要结果

记录了73组测量值。发现脉搏血氧测定法与动脉闭塞法之间的相关性最佳(r2 = 0.99)。间接测量方法之间的相关性比与直接动脉内测量的相关性更好。直升机的噪音和振动并未对脉搏血氧仪的操作产生明显干扰。

结论

我们得出结论,能够显示容积描记波形的脉搏血氧仪可在飞行中准确测量收缩压。