Caulkett N A, Duke T, Bailey J V
Department of Veterinary Anesthesiology, Radiology and Surgery, Western College of Veterinary Medicine, University of Saskatchewan, Saskatoon.
Can J Vet Res. 1994 Apr;58(2):144-7.
Systolic blood pressure measurement obtained with a pulse oximeter has been compared to values obtained by other indirect methods in man. Direct pressure measurement is subject to less error than indirect techniques. This study was designed to compare systolic pressure values obtained using a pulse oximeter, with values obtained by direct arterial pressure measurement. The pulse oximeter waveform was used as an indication of perfusion. A blood pressure cuff was applied proximal to the pulse oximeter probe. The cuff was inflated until the oximeter waveform disappeared, this value was recorded as the systolic pressure at the disappearance of the waveform (SPD). The cuff was inflated to a pressure > 200 mmHg, then gradually deflated until the waveform reappeared, this value was recorded as the systolic pressure at reappearance of the waveform (SPR). The average of the two values, SPD and SPR, was calculated and recorded as SPA. The study was performed in sows (n = 21) undergoing cesarean section under epidural anesthesia and IV sedation. A total of 280 measurements were made of SPD, SPR and SPA. Regression analysis of SPA and direct measurement revealed a correlation coefficient (r) of 0.81. Calculation of mean difference (bias) and standard deviation of the bias (precision) for direct pressure--SPA revealed a value of 1.3 +/- 12.1. When compared with direct measurement, the correlation of this technique was similar to that recorded for other indirect techniques used in small animals. This indicates that this technique would be useful for following systolic pressure trends.(ABSTRACT TRUNCATED AT 250 WORDS)
已将通过脉搏血氧仪获得的收缩压测量值与人类通过其他间接方法获得的值进行了比较。直接压力测量比间接技术的误差更小。本研究旨在比较使用脉搏血氧仪获得的收缩压值与通过直接动脉压测量获得的值。脉搏血氧仪波形用作灌注的指标。在脉搏血氧仪探头近端施加血压袖带。袖带充气直至血氧仪波形消失,该值记录为波形消失时的收缩压(SPD)。袖带充气至压力>200 mmHg,然后逐渐放气直至波形再次出现,该值记录为波形再次出现时的收缩压(SPR)。计算SPD和SPR这两个值的平均值并记录为SPA。该研究在硬膜外麻醉和静脉镇静下接受剖宫产的母猪(n = 21)中进行。共对SPD、SPR和SPA进行了280次测量。对SPA与直接测量值进行回归分析,得出相关系数(r)为0.81。直接压力 - SPA的平均差异(偏差)和偏差的标准差(精密度)计算得出的值为1.3±12.1。与直接测量相比,该技术的相关性与小动物中使用的其他间接技术记录的相关性相似。这表明该技术可用于追踪收缩压趋势。(摘要截短为250字)