Department of Radiation Oncology, University Medical Center Mannheim, University of Heidelberg, Mannheim, Germany.
Int J Radiat Oncol Biol Phys. 2010 Nov 15;78(4):1219-26. doi: 10.1016/j.ijrobp.2010.01.030. Epub 2010 May 28.
Hypofractionated high-dose radiotherapy for small lung tumors has typically been based on stereotaxy. Cone-beam computed tomography and breath-hold techniques have provided a noninvasive basis for precise cranial and extracranial patient positioning. The cone-beam computed tomography acquisition time of 60 s, however, is beyond the breath-hold capacity of patients, resulting in respiratory motion artifacts. By combining megavoltage (MV) and kilovoltage (kV) photon sources (mounted perpendicularly on the linear accelerator) and accelerating the gantry rotation to the allowed limit, the data acquisition time could be reduced to 15 s.
An Elekta Synergy 6-MV linear accelerator, with iViewGT as the MV- and XVI as the kV-imaging device, was used with a Catphan phantom and an anthropomorphic thorax phantom. Both image sources performed continuous image acquisition, passing an angle interval of 90° within 15 s. For reconstruction, filtered back projection on a graphics processor unit was used. It reconstructed 100 projections acquired to a 512 × 512 × 512 volume within 6 s.
The resolution in the Catphan phantom (CTP528 high-resolution module) was 3 lines/cm. The spatial accuracy was within 2-3 mm. The diameters of different tumor shapes in the thorax phantom were determined within an accuracy of 1.6 mm. The signal-to-noise ratio was 68% less than that with a 180°-kV scan. The dose generated to acquire the MV frames accumulated to 82.5 mGy, and the kV contribution was <6 mGy.
The present results have shown that fast breath-hold, on-line volume imaging with a linear accelerator using simultaneous kV-MV cone-beam computed tomography is promising and can potentially be used for image-guided radiotherapy for lung cancer patients in the near future.
小肺肿瘤的适形大剂量放疗通常基于立体定向技术。锥形束 CT 和屏气技术为精确的颅内外患者定位提供了非侵入性的基础。然而,锥形束 CT 的采集时间为 60 秒,超出了患者屏气的能力,导致呼吸运动伪影。通过组合兆伏(MV)和千伏(kV)光子源(垂直安装在直线加速器上)并将机架旋转加速到允许的极限,数据采集时间可以缩短到 15 秒。
使用 Elekta Synergy 6-MV 直线加速器,MV 成像设备为 iViewGT,kV 成像设备为 XVI,结合 Catphan 体模和人体胸部体模进行实验。两种图像源均进行连续图像采集,在 15 秒内通过 90°的角度间隔。对于重建,在图形处理器单元上使用滤波反投影。它在 6 秒内重建了 100 个采集到的 512×512×512 体素的投影。
Catphan 体模(CTP528 高分辨率模块)的分辨率为 3 线/cm。空间精度在 2-3mm 以内。胸部体模中不同肿瘤形状的直径测量精度为 1.6mm。与 180°kV 扫描相比,信噪比降低了 68%。采集 MV 帧的剂量累积为 82.5mGy,kV 贡献小于 6mGy。
目前的结果表明,使用直线加速器进行快速屏气、在线容积成像,同时进行 kV-MV 锥形束 CT 是有前途的,有望在不久的将来用于肺癌患者的图像引导放疗。