Department of Electrical Engineering, Stanford University, Stanford, California, USA.
Magn Reson Med. 2010 Oct;64(4):1068-77. doi: 10.1002/mrm.22502.
Temperature estimation in proton resonance frequency (PRF) shift MR thermometry requires a reference, or pretreatment, phase image that is subtracted from image phase during thermal treatment to yield a phase difference image proportional to temperature change. Referenceless thermometry methods derive a reference phase image from the treatment image itself by assuming that in the absence of a hot spot, the image phase can be accurately represented in a smooth (usually low order polynomial) basis. By masking the hot spot out of a least squares (ℓ(2)) regression, the reference phase image's coefficients on the polynomial basis are estimated and a reference image is derived by evaluating the polynomial inside the hot spot area. Referenceless methods are therefore insensitive to motion and bulk main field shifts, however, currently these methods require user interaction or sophisticated tracking to ensure that the hot spot is masked out of the polynomial regression. This article introduces an approach to reference PRF shift thermometry that uses reweighted ℓ(1) regression, a form of robust regression, to obtain background phase coefficients without hot spot tracking and masking. The method is compared to conventional referenceless thermometry, and demonstrated experimentally in monitoring HIFU heating in a phantom and canine prostate, as well as in a healthy human liver.
在质子共振频率 (PRF) 偏移磁共振测温中,需要一个参考(或预处理)相位图像,该图像在热治疗期间从图像相位中减去,以产生与温度变化成正比的相位差图像。无参考测温方法通过假设在没有热点的情况下,图像相位可以用平滑(通常是低阶多项式)基准确表示,从处理图像本身推导出参考相位图像。通过从最小二乘(ℓ(2))回归中掩蔽热点,可以估计参考相位图像的多项式基系数,并通过在热点区域内评估多项式来得出参考图像。因此,无参考方法对运动和主磁场偏移不敏感,但是,目前这些方法需要用户交互或复杂的跟踪来确保将热点从多项式回归中掩蔽出来。本文介绍了一种使用加权 ℓ(1)回归(一种稳健回归形式)的参考 PRF 偏移测温方法,无需热点跟踪和掩蔽即可获得背景相位系数。该方法与传统的无参考测温方法进行了比较,并在模拟和犬前列腺以及健康人肝脏中进行了 HIFU 加热监测的实验中进行了演示。