Department of Orthopaedic Surgery, Siriraj Hospital, Mahidol University, Bangkok, Thailand.
J Arthroplasty. 2011 Apr;26(3):504.e13-6. doi: 10.1016/j.arth.2010.03.028. Epub 2010 May 31.
Fracture of the neck of the femoral component after total hip arthroplasty is an infrequent complication. We present a report of 2 cases of trunnion fractures of fully porous-coated femoral stems (AML A Plus, DePuy International, Leeds, England), which had been implanted for 6 and 7 years after the index procedures. In both cases, failure emanated from a region where the trunnion had been intentionally grooved; and in both cases, the diameter of the trunnion was small, and evidence of corrosion contributing to the initiation of the failure was noted. Although this is the first report of trunnion fractures in this particular implant design, a close monitoring of prostheses with this specific trunnion should be considered. In addition, further analysis of clinical results and complications with this implant is warranted.
髋关节置换术后股骨部件颈骨折是一种罕见的并发症。我们报告了 2 例全多孔涂层股骨柄(AML A Plus,DePuy International,Leeds,England)的臼杯骨折,这些病例在指数手术后 6 年和 7 年分别出现了该并发症。在这两种情况下,失败都源于臼杯被故意开槽的区域;在这两种情况下,臼杯的直径都较小,并且注意到腐蚀会导致故障的发生。尽管这是在这种特殊植入物设计中首次报告臼杯骨折,但应考虑对具有这种特定臼杯的假体进行密切监测。此外,还需要进一步分析该植入物的临床结果和并发症。