Department of Radiology, Stanford University, Stanford, California, USA.
J Magn Reson Imaging. 2010 Jul;32(1):173-83. doi: 10.1002/jmri.22213.
To compare six new three-dimensional (3D) magnetic resonance (MR) methods for evaluating knee cartilage at 3.0T.
We compared: fast-spin-echo cube (FSE-Cube), vastly undersampled isotropic projection reconstruction balanced steady-state free precession (VIPR-bSSFP), iterative decomposition of water and fat with echo asymmetry and least-squares estimation combined with spoiled gradient echo (IDEAL-SPGR) and gradient echo (IDEAL-GRASS), multiecho in steady-state acquisition (MENSA), and coherent oscillatory state acquisition for manipulation of image contrast (COSMIC). Five-minute sequences were performed twice on 10 healthy volunteers and once on five osteoarthritis (OA) patients. Signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were measured from the volunteers. Images of the five volunteers and the five OA patients were ranked on tissue contrast, articular surface clarity, reformat quality, and lesion conspicuity. FSE-Cube and VIPR-bSSFP were compared to IDEAL-SPGR for cartilage volume measurements.
FSE-Cube had top rankings for lesion conspicuity, overall SNR, and CNR (P < 0.02). VIPR-bSSFP had top rankings in tissue contrast and articular surface clarity. VIPR and FSE-Cube tied for best in reformatting ability. FSE-Cube and VIPR-bSSFP compared favorably to IDEAL-SPGR in accuracy and precision of cartilage volume measurements.
FSE-Cube and VIPR-bSSFP produce high image quality with accurate volume measurement of knee cartilage.
比较 6 种新的 3.0T 膝关节软骨 3D 磁共振(MR)成像方法。
我们比较了快速自旋回波体部采集(FSE-Cube)、大视野完全采样各向同性容积内插预备平衡快速梯度回波(VIPR-bSSFP)、水脂分离迭代分解双稳态自由进动(IDEAL-SPGR)和梯度回波(IDEAL-GRASS)、稳态采集多回波(MENSA)和相干振荡状态采集用于图像对比度操控(COSMIC)。在 10 名健康志愿者上分别进行 2 次 5 分钟序列扫描,5 例骨关节炎(OA)患者进行 1 次扫描。志愿者的信号噪声比(SNR)和对比噪声比(CNR)进行测量。5 名志愿者和 5 例 OA 患者的图像进行组织对比度、关节面清晰度、重建质量和病变显示程度的评分。FSE-Cube 和 VIPR-bSSFP 与 IDEAL-SPGR 行软骨体积测量比较。
FSE-Cube 在病变显示程度、总体 SNR 和 CNR 方面具有最高评分(P < 0.02)。VIPR-bSSFP 在组织对比度和关节面清晰度方面具有最高评分。VIPR 和 FSE-Cube 在重建能力方面得分相同。FSE-Cube 和 VIPR-bSSFP 在软骨体积测量的准确性和精确性方面与 IDEAL-SPGR 相当。
FSE-Cube 和 VIPR-bSSFP 可获得高质量图像,并可准确测量膝关节软骨体积。