Suppr超能文献

地氟醚对呼吸回路过滤器过滤性能的影响。

The effect of desflurane on filtration performance of breathing system filters.

机构信息

Department of Anaesthetics, University Hospital Birmingham, Edgbaston, Birmingham, UK.

出版信息

Anaesthesia. 2010 Aug;65(8):831-5. doi: 10.1111/j.1365-2044.2010.06419.x. Epub 2010 Jun 28.

Abstract

This study assessed the effect of desflurane on the filtration performance of six breathing system filters intended for use with adults. Three filters contained an electrostatic filter material and three contained a pleated glass fibre filter material. Five samples of each model of filter were exposed to 6% v/v of desflurane for 1 h, 12% v/v of desflurane for 1 h, 12% v/v of desflurane for 4 h and air only for 1 h. Five samples of each filter were also tested without exposure to any vapour or air. The filtration performance was measured by challenging each filter with an aerosol of sodium chloride particles using a Moore's test rig. Penetration of particles through the electrostatic filters increased following exposure to a higher concentration of desflurane for a longer duration (p < 0.001). The effect on two of the pleated filters was not significant (p = 0.55 and p = 0.64). The effect on the remaining pleated filter was significant (p < 0.001) but small. The efficiency of some filters decreases when they are exposed to high concentrations of desflurane for a long duration. This effect appears more marked in electrostatic filters compared with pleated filters.

摘要

本研究评估了地氟醚对六种成人用呼吸回路过滤器过滤性能的影响。三种过滤器含有静电滤料,三种含有褶皱玻璃纤维滤料。每种型号的过滤器各取 5 个样本,分别暴露于 6%v/v 的地氟醚 1 小时、12%v/v 的地氟醚 1 小时、12%v/v 的地氟醚 4 小时以及仅暴露于空气 1 小时。每种过滤器的 5 个样本也未暴露于任何蒸气或空气的情况下进行了测试。使用 Moore 测试台,通过用氯化钠颗粒气溶胶对每个过滤器进行挑战来测量过滤性能。静电过滤器在暴露于更高浓度的地氟醚更长时间后,颗粒穿透率增加(p < 0.001)。对其中两种褶皱过滤器的影响不显著(p = 0.55 和 p = 0.64)。对其余褶皱过滤器的影响显著(p < 0.001)但较小。当过滤器长时间暴露于高浓度的地氟醚时,其效率会降低。与褶皱过滤器相比,这种效应在静电过滤器中更为明显。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验