Eur Heart J. 2010 Sep;31(18):2246-52. doi: 10.1093/eurheartj/ehq203. Epub 2010 Jun 30.
Automatic daily transmission of data from implantable cardioverter-defibrillators (ICDs) enables the remote monitoring of device status and leads function. We report on a 2-year experience with remote monitoring in 40 recipients of high-voltage ICD leads, prone to fracture and under advisory since October 2007.
The ICDs were remotely monitored as well as systematically interrogated in the ambulatory department every 3 months. The patients were also seen in case of abnormal lead impedance, or other manifestations consistent with lead dysfunction. Over a mean follow-up of 22 ± 4 months after ICD implantation, four lead dysfunctions were suspected because of remotely transmitted oversensing of noise artifacts, abrupt rise in pacing impedance, or both. A lead fracture needing lead replacement was confirmed in three patients (7.5%), two of them before any inappropriate therapy and one after the delivery of three inappropriate shocks. No lead failure was observed in the remaining 36 patients, either at the time of ambulatory visits or during remote monitoring.
Remote monitoring allowed the early and reliable detection of ICD leads failure without requiring any patient intervention.
植入式心脏复律除颤器(ICD)的自动日常数据传输可实现设备状态和功能的远程监测。我们报告了自 2007 年 10 月以来,对 40 例易发生断裂且有警示作用的高压 ICD 导联患者进行为期 2 年的远程监测的经验。
对 ICD 进行远程监测,并在门诊每 3 个月进行系统询问。当出现异常导联阻抗或其他与导联功能障碍一致的表现时,患者也会就诊。在 ICD 植入后平均 22±4 个月的随访期间,由于远程传输的噪声伪影过感知、起搏阻抗突然升高或两者兼而有之,怀疑有 4 例导联功能障碍。在 3 名患者(7.5%)中确认了需要更换导联的导联断裂,其中 2 例在任何不适当治疗之前,1 例在发生 3 次不适当电击后。在其余 36 名患者中,无论是在门诊就诊时还是在远程监测期间,均未观察到导联故障。
远程监测可早期、可靠地检测 ICD 导联故障,而无需任何患者干预。