Department of Pathology and Laboratory Medicine, University of Texas Medical School at Houston, Houston, Texas, USA.
J Clin Lab Anal. 2010;24(4):278-82. doi: 10.1002/jcla.20400.
Carbamazepine is an anticonvulsant requiring routine therapeutic drug monitoring. Recently, Siemens Healthcare Diagnostic Division released a new carbamazepine assay: ADVIA Chemistry Carbamazepine_2 (Carbamazepine_2) for application on ADVIA analyzers. We evaluated the analytical performance of this assay as well as its potential cross-reactivities with carbamazepine 10, 11-epoxide, hydroxyzine, and cetirizine. The within-run and between-run precisions of the Carbamzepine-2 assay were <6% and limit of detection was 0.5 microg/ml using ADVIA 1800 analyzer. The assay was linear up to a carbamazepine concentration of 20.0 microg/ml. The new method compared well with a widely used carbamazepine EMIT 2000 assay on the Hitachi 917 analyzer. Using 75 patients' specimens (where carbamazepine concentrations varied from 0.5 to 21.7 microg/ml) and carbamazepine EMIT 2000 as the reference method (x-axis), we observed the following regression equation: y=1.04 x+0.32 (r=0.99). The new carbazepine_2 method was not affected by a hemoglobin concentration of 1,000 mg/dl, conjugated or unconjugated bilirubin concentration of 60 mg/dl, and triglyceride concentration of 1,000 mg/dl. In addition, this assay showed no cross-reactivity with hydroxyzine or cetirizine and demonstrated minimal cross-reactivity with carbamazepine 10, 11-epoxide. We conclude that the ADVIA Chemistry carbamazepine_2 assay has adequate precision and accuracy for routine therapeutic drug monitoring of carbamazepine in clinical laboratories.
卡马西平是一种抗惊厥药物,需要常规进行治疗药物监测。最近,西门子医疗诊断部门发布了一种新的卡马西平检测方法:ADVIA Chemistry Carbamazepine_2(卡马西平_2),适用于 ADVIA 分析仪。我们评估了该检测方法的分析性能及其与卡马西平 10,11-环氧物、羟嗪和西替利嗪的潜在交叉反应性。使用 ADVIA 1800 分析仪,Carbamzepine-2 检测的批内和批间精密度均<6%,检测限为 0.5μg/ml。该检测方法在卡马西平浓度为 20.0μg/ml 时呈线性。新方法与日立 917 分析仪上广泛使用的卡马西平 EMIT 2000 检测方法相比表现良好。使用 75 名患者的标本(卡马西平浓度从 0.5 到 21.7μg/ml 不等)和卡马西平 EMIT 2000 作为参考方法(x 轴),我们观察到以下回归方程:y=1.04x+0.32(r=0.99)。新的卡马西平_2 方法不受血红蛋白浓度为 1000mg/dl、结合或未结合胆红素浓度为 60mg/dl 和甘油三酯浓度为 1000mg/dl 的影响。此外,该检测方法与羟嗪或西替利嗪无交叉反应性,与卡马西平 10,11-环氧物的交叉反应性最小。我们得出结论,ADVIA Chemistry 卡马西平_2 检测方法具有足够的精密度和准确性,可用于临床实验室中卡马西平的常规治疗药物监测。