Suppr超能文献

由于存在羟嗪和西替利嗪,使用荧光偏振免疫分析法会出现血清三环类抗抑郁药浓度假阳性。

False-positive serum tricyclic antidepressant concentrations using fluorescence polarization immunoassay due to the presence of hydroxyzine and cetirizine.

作者信息

Dasgupta Amitava, Wells Alice, Datta Pradip

机构信息

Department of Pathology and Laboratory Medicine, University of Texas Medical School at Houston, Houston, Texas 77030, USA.

出版信息

Ther Drug Monit. 2007 Feb;29(1):134-9.

Abstract

A recent report indicates that hydroxyzine and its active metabolite cetirizine interfere with the PENTINA carbamazepine assay. The potential interference of hydroxyzine and cetirizine with the fluorescence polarization immunoassay (FPIA) and CEDIA assay of carbamazepine as well as with the fluorescence polarization immunoassay of tricyclic antidepressants (TCA) was studied. Aliquots of drug-free serum pools were supplemented with various concentrations of hydroxyzine and cetirizine representing therapeutic, mild to moderate toxic as well as very toxic concentrations. Then apparent carbamazepine and TCA concentrations were measured by immunoassays. Although no interference of hydroxyzine and cetirizine was observed with carbamazepine assays (FPIA and CEDIA), significant apparent TCA concentrations were observed when aliquots of drug-free serum were supplemented with hydroxyzine or cetirizine. Mathematical formula was devised to predict hydroxyzine and/or cetirizine concentration in serum based on observed apparent TCA levels. Hydroxyzine and cetirizine also falsely increased total TCA values when aliquots of serum pool prepared from patients receiving TCA were further supplemented with these drugs. In conclusion, hydroxyzine and cetirizine do not interfere with the FPIA and CEDIA carbamazepine assays but interfere with the measurement of total TCA using the FPIA.

摘要

最近一份报告指出,羟嗪及其活性代谢物西替利嗪会干扰卡马西平的PENTINA检测法。研究了羟嗪和西替利嗪对卡马西平荧光偏振免疫分析(FPIA)和CEDIA分析以及三环类抗抑郁药(TCA)荧光偏振免疫分析的潜在干扰。向不含药物的血清混合液等分试样中添加不同浓度的羟嗪和西替利嗪,这些浓度代表治疗浓度、轻度至中度中毒浓度以及极毒浓度。然后通过免疫分析法测量卡马西平和TCA的表观浓度。尽管未观察到羟嗪和西替利嗪对卡马西平检测法(FPIA和CEDIA)有干扰,但当向不含药物的血清等分试样中添加羟嗪或西替利嗪时,观察到了显著的表观TCA浓度。设计了数学公式,根据观察到的表观TCA水平预测血清中羟嗪和/或西替利嗪的浓度。当向接受TCA治疗的患者制备的血清混合液等分试样中进一步添加这些药物时,羟嗪和西替利嗪还会错误地增加总TCA值。总之,羟嗪和西替利嗪不干扰FPIA和CEDIA卡马西平检测法,但会干扰使用FPIA测量总TCA。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验