How does an acquisition at reduced doses using automatic tube current modulation techniques compare to the normal standard dose CT? Does it affect the sensitivity for detection of calcifications? CT raw data of 54 patients with suspected urolithiasis acquired with automatic tube current modulation techniques were used for image noise simulations with 100%, 50% and 25% dose simulated. Data were analyzed by independent readers with regard to the presence of urolithiasis, stone location, size, density and differential diagnoses. The mean effective dose per standard examination/50%/25% simulation was 7.3 mSv/3.8 mSv/1.9 mSv. Sensitivities/specificities for detection of urolithiasis were calculated for all dose simulations and resulted in 0.94/0.98 in the 50% dose level group and 0.82/0.97 in the 25% dose level group. Low-dose CT with tube current modulation can be used as a standard procedure for the evaluation of patients with suspected acute renal colic.
使用自动管电流调制技术降低剂量进行采集与常规标准剂量 CT 相比如何?它会影响对钙化检测的敏感性吗?使用自动管电流调制技术对 54 例疑似尿路结石的患者进行 CT 采集,获得 CT 原始数据,模拟 100%、50%和 25%剂量的图像噪声。由独立的读者对尿路结石的存在、结石位置、大小、密度和鉴别诊断进行分析。每次标准检查/50%/25%模拟的平均有效剂量分别为 7.3 mSv/3.8 mSv/1.9 mSv。对所有剂量模拟计算结石检测的敏感性/特异性,在 50%剂量组分别为 0.94/0.98,在 25%剂量组为 0.82/0.97。管电流调制的低剂量 CT 可作为评估疑似急性肾绞痛患者的标准程序。