Comparison of exhalation time methods (6 sec vs. 10 sec) of a hand-held exhaled nitric oxide analyzer.

机构信息

Department of Pediatrics, Faculty of Medicine, University of Toyama, Toyama, Japan.

出版信息

Pediatr Pulmonol. 2010 Oct;45(10):1005-8. doi: 10.1002/ppul.21286.

Abstract

Standard exhalation time for measuring fractional exhaled nitric oxide (FeNO) is 10 sec, but this is not easy for younger children. We aimed to investigate the agreement between FeNO values during 10-sec (FeNO-10) and 6-sec (FeNO-6) exhalation and the feasibility of measuring FeNO-6, using a hand-held analyzer, NIOX-MINO®. FeNO values measured during 10- and 6-sec (random order) were compared. Success rates of the two different time modes were also evaluated. In 119 asthmatic children (median age 8 years [range 4-15]) who had been already accustomed to NIOX-MINO®, median FeNO-10 (29 ppb [IQR 15.2-42.0]) and FeNO-6 (27 ppb [IQR 16.0-43.5]) did not differ significantly (P = 0.90), and there was a good correlation between both values (r = 0.984, P < 0.001). Mean difference (FeNO-10-FeNO-6) was -0.151 ppb (95% CI: -0.95 to 0.65, limits of agreement: -8.8 to 8.5). In 46 asthmatic children (median age 7 years [range 4-15]) who had never used any FeNO analyzers, all the children aged 8 years and more (n = 21) succeeded in measuring FeNO on both time modes, whereas for children aged younger than 8 years (n = 25) success rates of the 10- and 6-sec mode were 60.0% and 92.0%, respectively. In conclusion, we showed good agreement between FeNO-10 and FeNO-6, and the 6-sec mode of NIOX-MINO® is more feasible than 10-sec mode for measuring FeNO in younger children.

摘要

标准的呼出气一氧化氮(FeNO)测量时间为 10 秒,但这对于年幼的儿童来说并不容易。我们旨在研究使用手持式分析仪 NIOX-MINO®测量 10 秒(FeNO-10)和 6 秒(FeNO-6)呼气期间 FeNO 值之间的一致性,以及测量 FeNO-6 的可行性。比较了 10 秒和 6 秒(随机顺序)期间测量的 FeNO 值。还评估了两种不同时间模式的成功率。在已经习惯使用 NIOX-MINO®的 119 名哮喘儿童(中位数年龄为 8 岁[范围 4-15])中,FeNO-10(29 ppb [IQR 15.2-42.0])和 FeNO-6(27 ppb [IQR 16.0-43.5])中位数差异无统计学意义(P = 0.90),并且两者之间存在良好的相关性(r = 0.984,P < 0.001)。平均差异(FeNO-10-FeNO-6)为-0.151 ppb(95%CI:-0.95 至 0.65,一致性界限:-8.8 至 8.5)。在从未使用过任何 FeNO 分析仪的 46 名哮喘儿童(中位数年龄为 7 岁[范围 4-15])中,所有 8 岁及以上的儿童(n = 21)均成功在两种时间模式下测量了 FeNO,而对于年龄小于 8 岁的儿童(n = 25),10 秒和 6 秒模式的成功率分别为 60.0%和 92.0%。总之,我们发现 FeNO-10 和 FeNO-6 之间具有良好的一致性,并且对于测量年幼儿童的 FeNO 来说,NIOX-MINO®的 6 秒模式比 10 秒模式更可行。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验