The objective of this study was to compare the LCh° color coordinates of dental color-measuring devices with those of a spectrophotometric reference system.
The International Commission on Illumination (CIE) LCh° color coordinates of ceramic samples, matching the colors of the VITA Linearguide, were recorded using four color-measuring devices (VITA Easyshade (A), VITA Easyshade compact (B), Degudent Shadepilot (C), X-Rite Shadevision (D)) and a spectrophotometric reference system under standardized test conditions. The intraclass correlation coefficients, regression lines with coefficients of determination, and mean deviations of the dental color-measuring devices from the spectrophotometric reference system were calculated.
All of the devices had high intraclass correlation coefficients, between 0.979 and 1.000. Nearly all measurements, except the hue values of devices A, B, and D, showed coefficients of determination close to 1.0 over the range of measurements for LCh° color coordinates. Similarly, all of the devices had coefficients of determination near the optimum value of 1.0. Devices C and D produced regression line slopes near the optimum value of 1.0 and intercepts close to the optimum value of zero. Only the L* coordinate measurements of devices A and B and the h° coordinate measurements of device B differed not significantly from the spectrophotometric reference values.
The electronic dental color-measuring devices tested showed excellent repeatability, but some devices showed substantial deviations in color coordinate values from the spectrophotometric reference system. Devices C and D showed higher precision than devices A and B.
本研究的目的是比较牙科比色仪的 LCh°颜色坐标与分光光度参考系统的颜色坐标。
使用四种比色仪(VITA Easyshade(A)、VITA Easyshade compact(B)、Degudent Shadepilot(C)、X-Rite Shadevision(D))和分光光度参考系统,记录与 VITA Linearguide 颜色匹配的陶瓷样本的国际照明委员会(CIE)LCh°颜色坐标,在标准化测试条件下进行。计算了牙科比色仪的组内相关系数、决定系数的回归线和与分光光度参考系统的平均偏差。
所有仪器的组内相关系数均较高,在 0.979 到 1.000 之间。除了仪器 A、B 和 D 的色调值外,几乎所有的测量值都表现出接近 1.0 的决定系数,在 LCh°颜色坐标的测量范围内。同样,所有仪器的决定系数都接近最佳值 1.0。仪器 C 和 D 的回归线斜率接近最佳值 1.0,截距接近最佳值 0。只有仪器 A 和 B 的 L*坐标测量值和仪器 B 的 h°坐标测量值与分光光度参考值没有显著差异。
所测试的电子牙科比色仪具有出色的重复性,但在颜色坐标值方面,一些仪器与分光光度参考系统存在较大偏差。仪器 C 和 D 比仪器 A 和 B 具有更高的精度。