与颌骨特征及种植体植入早期愈合过程相关的共振频率分析

Resonance frequency analysis in relation to jawbone characteristics and during early healing of implant installation.

作者信息

Huwiler M A, Pjetursson B E, Bosshardt D D, Salvi G E, Lang N P

机构信息

School of Dental Medicine, University of Berne, Berne, Switzerland.

出版信息

Clin Oral Implants Res. 2007 Jun;18(3):275-80. doi: 10.1111/j.1600-0501.2007.01336.x. Epub 2007 Mar 12.

Abstract

OBJECTIVES

To monitor resonance frequency analysis (RFA) in relation to the jawbone characteristics and during the early phases of healing and incorporation of Straumann dental implants with an SLA surface.

MATERIAL AND METHODS

17 Straumann 4.1 mm implants (10 mm) and 7 Straumann 4.8 mm implants (10 mm) were installed and ISQ determined at baseline and after 1, 2, 3, 4, 5, 6, 8 and 12 weeks. Central bone cores were analyzed from the 4.1 mm implants using micro CT for bone volume density (BVD) and bone trabecular connectivity (BTC).

RESULTS

Pocket probing depths ranged from 2-4 mm and bleeding on probing from 5-20%. At baseline, BVD varied between 24% and 65% and BTC between 4.9 and 25.4 for the 4.1 mm implants. Baseline ISQ varied between 55 and 74 with a mean of 61.4. No significant correlations were found between BVD or BTC and ISQ Values. For the 4.8 mm diameter implants baseline ISQ values ranged from 57-70 with a mean of 63.3. Over the healing period ISQ values increased at 1 week and decreased after 2-3 weeks. After 4 weeks ISQ values, again increased slightly, no significant differences were noted over time. One implant (4.1 mm) lost stability at 3 weeks. Its ISQ value had dropped from 68 to 45. However the latter value was determined after the clinical diagnosis of instability.

CONCLUSION

ISQ values of 57-70 represented homeostasis and implant stability. However no predictive value for loosing implant stability can be attributed to RFA since the decrease occurred after the fact.

摘要

目的

监测共振频率分析(RFA)与颌骨特征的关系,以及在具有SLA表面的士卓曼牙科种植体愈合和整合的早期阶段的情况。

材料与方法

植入17颗士卓曼4.1毫米种植体(10毫米)和7颗士卓曼4.8毫米种植体(10毫米),并在基线以及1、2、3、4、5、6、8和12周后测定植入稳定性商数(ISQ)。使用微型计算机断层扫描(micro CT)分析4.1毫米种植体的中央骨芯,以测定骨体积密度(BVD)和骨小梁连通性(BTC)。

结果

牙周袋探诊深度为2 - 4毫米,探诊出血为5 - 20%。在基线时,4.1毫米种植体的BVD在24%至65%之间变化,BTC在4.9至25.4之间变化。基线ISQ在55至74之间变化,平均值为61.4。未发现BVD或BTC与ISQ值之间存在显著相关性。对于4.8毫米直径的种植体,基线ISQ值在57 - 70之间,平均值为63.3。在愈合期,ISQ值在1周时增加,在2 - 3周后下降。4周后,ISQ值再次略有增加,随时间未观察到显著差异。一颗种植体(4.1毫米)在3周时失去稳定性。其ISQ值从68降至45。然而,后者的值是在临床诊断为不稳定后测定的。

结论

ISQ值在57 - 70表示处于稳态和种植体稳定。然而,由于种植体稳定性下降是在实际发生后才出现的,因此共振频率分析(RFA)对于预测种植体稳定性丧失没有预测价值。