应用临床普尔金耶(Purkinje)目镜测量人工晶状体偏心和倾斜的重复性。

Reproducibility of intraocular lens decentration and tilt measurement using a clinical Purkinje meter.

机构信息

Department of Pathology, Institute of Ophthalmology, University College London, and Moorfields Eye Hospital NHS Foundation Trust, London, United Kingdom.

出版信息

J Cataract Refract Surg. 2010 Sep;36(9):1529-35. doi: 10.1016/j.jcrs.2010.03.043.

Abstract

PURPOSE

To determine the reproducibility of intraocular lens (IOL) decentration and tilt measurements with a new Purkinje meter instrument.

SETTING

Moorfields Eye Hospital NHS Foundation Trust, London, United Kingdom.

METHODS

After pupil dilation, images of pseudophakic eyes with a plate-style IOL (Akreos Adapt) were obtained using a recently developed Purkinje meter. Intraocular lens decentration and tilt were evaluated by analyzing the captured images using a semiobjective method by marking the reflexes in the images and automatic calculation using a dedicated software program. In study 1, examiner 1 examined the eyes first followed by examiner 2. Ten minutes later, examiner 1 performed a second measurement, after which the intraexaminer and interexaminer reproducibility were determined. In study 2, a Purkinje meter was used to measure pseudophakic eyes with slitlamp finding of clinical IOL decentration, IOL tilt, or both. The results were compared with retroillumination photographs and slitlamp findings.

RESULTS

In study 1, there was high intraexaminer reproducibility for decentration (r = 0.95) and tilt (r = 0.85) and high interexaminer reproducibility for decentration (r = 0.84) and tilt (r = 0.75). In study 2, even in extreme cases of decentration and/or tilt, the Purkinje meter measurements were possible and appeared to correlate well with slitlamp findings.

CONCLUSIONS

Acquisition of images in pseudophakic eyes with the Purkinje meter was simple and rapid. The method was highly reliable for 1 examiner and between 2 examiners.

摘要

目的

确定新型普尔金耶(Purkinje)米计仪器测量人工晶状体(IOL)偏心和倾斜的可重复性。

设置

英国伦敦 Moorfields Eye Hospital NHS 基金会信托。

方法

瞳孔扩张后,使用最近开发的普尔金耶米计获取具有板式 IOL(Akreos Adapt)的人工晶状体眼的像。通过分析捕获的图像,使用半客观方法评估人工晶状体偏心和倾斜,在图像中标记反射,并使用专用软件程序自动计算。在研究 1 中,检查者 1 首先检查眼睛,然后检查者 2 检查。10 分钟后,检查者 1 进行第二次测量,然后确定检查者内和检查者间的可重复性。在研究 2 中,使用普尔金耶米计测量临床发现 IOL 偏心、IOL 倾斜或两者都有的人工晶状体眼。结果与背面照明照片和裂隙灯检查结果进行比较。

结果

在研究 1 中,偏心(r = 0.95)和倾斜(r = 0.85)的检查者内可重复性高,偏心(r = 0.84)和倾斜(r = 0.75)的检查者间可重复性高。在研究 2 中,即使在偏心和/或倾斜的极端情况下,普尔金耶米计的测量也是可行的,并且似乎与裂隙灯检查结果相关性良好。

结论

在人工晶状体眼采集普尔金耶米计图像简单且快速。该方法对于 1 名检查者和 2 名检查者均高度可靠。