Tilt and decentration of three-piece foldable high-refractive silicone and hydrophobic acrylic intraocular lenses with 6-mm optics in an intraindividual comparison.

作者信息

Baumeister Martin, Neidhardt Barbara, Strobel Jürgen, Kohnen Thomas

机构信息

Department of Ophthalmology, Johann Wolfgang Goethe University, Theodor-Stern-Kai 7, 60590 Frankfurt am Main, Germany.

出版信息

Am J Ophthalmol. 2005 Dec;140(6):1051-8. doi: 10.1016/j.ajo.2005.07.026.

Abstract

PURPOSE

Intraindividual comparison of tilt and decentration of three-piece foldable intraocular lenses (IOLs) with 6-mm optics and different edge design and material.

DESIGN

Prospective randomized study.

METHODS

Twenty-five patients with senile cataract (group I) received a foldable silicone, sharp optic edge IOL in one eye and a silicone, rounded optic edge IOL in the other eye. Group II (n = 28) received the foldable silicone, sharp optic edge IOL in one eye and an acrylate, sharp optic edge IOL in the other eye. Scheimpflug photography was performed after the procedure with an anterior eye segment analysis system. Tilt and decentration of the IOL optic were measured 1 week, 6 months, and 12 months after the procedure.

RESULTS

In group I, the foldable silicone, sharp optic edge IOL showed a mean optic tilt of 3.03 +/- 1.79 degrees and an optic decentration of 0.24 +/- 0.13 mm; the silicone, rounded optic edge IOL showed a tilt of 3.26 +/- 1.69 degrees and a decentration of 0.23 +/- 0.13 mm. In group II, the foldable silicone, sharp optic edge IOL showed an average tilt of 2.34 +/- 1.81 degrees and a decentration of 0.29 +/- 0.21 mm after 12 months; the acrylate, sharp optic edge IOL had a tilt of 2.32 +/- 1.41 degrees and a decentration of 0.24 +/- 0.10 mm. There were no significant differences in either group.

CONCLUSION

The examined IOLs showed a stable position regarding tilt and decentration in the first 12 postoperative months, independent of material and edge design. Slight deviations from the optical axis might affect the outcome in aspheric or multifocal IOLs.

摘要

目的

对三片式可折叠人工晶状体(IOL)的倾斜和偏心进行个体内比较,这些人工晶状体光学部直径为6mm,边缘设计和材料不同。

设计

前瞻性随机研究。

方法

25例老年性白内障患者(I组)一只眼植入可折叠硅凝胶材质、光学边缘锐利的人工晶状体,另一只眼植入硅凝胶材质、光学边缘圆润的人工晶状体。II组(n = 28)一只眼植入可折叠硅凝胶材质、光学边缘锐利的人工晶状体,另一只眼植入丙烯酸酯材质、光学边缘锐利的人工晶状体。术后使用眼前节分析系统进行Scheimpflug摄影。在术后1周、6个月和12个月测量人工晶状体光学部的倾斜和偏心。

结果

I组中,可折叠硅凝胶材质、光学边缘锐利的人工晶状体平均光学倾斜度为3.03±1.79度,光学偏心度为0.24±0.13mm;硅凝胶材质、光学边缘圆润的人工晶状体倾斜度为3.26±1.69度,偏心度为0.23±0.13mm。II组中,可折叠硅凝胶材质、光学边缘锐利的人工晶状体在12个月后平均倾斜度为2.34±1.81度,偏心度为0.29±0.21mm;丙烯酸酯材质、光学边缘锐利的人工晶状体倾斜度为2.32±1.41度,偏心度为0.24±0.10mm。两组均无显著差异。

结论

在所检查的人工晶状体中,术后12个月内,人工晶状体在倾斜和偏心方面位置稳定,与材料和边缘设计无关。与光轴的轻微偏差可能会影响非球面或多焦点人工晶状体的效果。