Suppr超能文献

用于小射野光子束的商用电离室中的极性效应。

Polarity effect in commercial ionization chambers used in photon beams with small fields.

作者信息

Shimono Tetsunori, Koshida Kichiro, Nambu Hidekazu, Matsubara Kosuke, Takahashi Hiroshi, Okuda Hiroshi

机构信息

Department of Radiology, Hoshigaoka Koseinenkin Hospital, Hirakata, Osaka, Japan.

出版信息

Radiol Phys Technol. 2009 Jan;2(1):97-103. doi: 10.1007/s12194-008-0050-1. Epub 2008 Dec 18.

Abstract

Ionization chambers are the instruments of choice for use in photon dosimetry. Ionization chambers together with radiographic films represent the best detectors for measurement of dose distribution for a quality assurance (QA) program in intensity-modulated radiotheraphy (IMRT). The polarity effect was investigated for seven different commercially available ionization chambers. This study concentrated on measuring the dependence of the polarity effect at various energies, and for various field size, ionization chamber, and electrometer combinations. Of the seven chambers, CC01, PTW23323, and PTW31006 had the largest polarity effect for small field sizes. The materials of the central electrode of these three chambers were steel or air-equivalent plastic C-552. The magnitude of the polarity effect was shown to be strongly dependent on the material of the collecting electrode. This polarity effect dependence was observed for the ionization chambers and small field sizes studied.

摘要

电离室是用于光子剂量测定的首选仪器。电离室与射线照相胶片一起,是调强放射治疗(IMRT)质量保证(QA)计划中剂量分布测量的最佳探测器。对七种不同的市售电离室的极性效应进行了研究。本研究集中于测量不同能量、不同射野大小、不同电离室和静电计组合下极性效应的依赖性。在这七个电离室中,CC01、PTW23323和PTW31006在小射野时具有最大的极性效应。这三个电离室中心电极的材料是钢或空气等效塑料C-552。结果表明,极性效应的大小强烈依赖于收集电极的材料。在所研究的电离室和小射野中观察到了这种极性效应依赖性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验