Department of Anesthesiology, University of Illinois at Chicago, Chicago, IL 60612, USA.
J Neurosurg Anesthesiol. 2010 Oct;22(4):309-15. doi: 10.1097/ANA.0b013e3181e4b7c8.
In earlier studies, nitrous oxide (N2O) did not affect bispectral index (BIS) or state entropy (SE) when administered as the sole anesthetic agent. We investigated the effect of adding N2O to sevoflurane or propofol anesthesia on BIS and SE.
A total of 28 patients were randomized to receive general anesthesia using either 2% sevoflurane or 120 mcg/kg/min propofol during surgery. After 20 minutes of stable anesthetic and surgical conditions (baseline), 60% N2O was added for 20 minutes. Another 20 minutes was allowed for washout after N2O was discontinued. Sevoflurane concentration was closely monitored using the standard end-tidal gas analysis and vaporizer output was constantly adjusted to offset the second-gas effect caused by the addition and discontinuation of N2O. Mean values for BIS and SE were calculated for the last 5 minutes of the baseline, nitrous oxide, and washout periods.
In the propofol group, BIS and SE values in baseline versus N2O periods were 40±10 versus 39±10 and 43±11 versus 43±12, respectively (mean±SD, P>0.05). In the sevoflurane group, addition of N2O decreased both BIS (36±7 to 32±7, P<0.001) and SE (37±10 to 31±11, P=0.001). The magnitude of change from the baseline in BIS and SE in the sevoflurane group was 3.6±3.8 (95% CI: 0.8-6.5) and 6.1±5.5 (95% CI: 3.7-8.6), respectively. Average baseline and washout values of BIS and SE were not different in both groups (P>0.05).
N2O decreased both BIS and SE when added to sevoflurane, but not propofol. The observed changes in the sevoflurane group were not clinically significant. Decreases in BIS and SE in the sevoflurane group could result from a true additive effect and second-gas effect of N2O that was unaccounted for despite a meticulous titration of sevoflurane using end-tidal gas monitoring.
在早期的研究中,一氧化二氮(N2O)作为唯一的麻醉剂时,不会影响脑电双频指数(BIS)或状态熵(SE)。我们研究了在七氟醚或异丙酚麻醉中加入 N2O 对 BIS 和 SE 的影响。
共有 28 名患者随机接受手术中使用 2%七氟醚或 120 mcg/kg/min 异丙酚的全身麻醉。在稳定的麻醉和手术条件 20 分钟后(基线),加入 60%的 N2O 20 分钟。停止 N2O 后,再给予 20 分钟洗脱期。使用标准呼气末气体分析密切监测七氟醚浓度,并不断调整挥发罐输出以抵消 N2O 加入和停止引起的第二气体效应。在基线、N2O 和洗脱期的最后 5 分钟计算 BIS 和 SE 的平均值。
在异丙酚组中,基线与 N2O 期的 BIS 和 SE 值分别为 40±10 比 39±10 和 43±11 比 43±12(平均值±标准差,P>0.05)。在七氟醚组中,加入 N2O 降低了 BIS(36±7 至 32±7,P<0.001)和 SE(37±10 至 31±11,P=0.001)。七氟醚组 BIS 和 SE 从基线的变化幅度分别为 3.6±3.8(95%CI:0.8-6.5)和 6.1±5.5(95%CI:3.7-8.6)。两组的 BIS 和 SE 的平均基线和洗脱值均无差异(P>0.05)。
当添加到七氟醚中时,N2O 降低了 BIS 和 SE,但不降低异丙酚。七氟醚组观察到的变化无临床意义。七氟醚组 BIS 和 SE 的降低可能是由于 N2O 的真实附加效应和第二气体效应所致,尽管使用呼气末气体监测对七氟醚进行了精细滴定,但仍未考虑到这一点。