The J Vernon Luck Sr, MD, Orthopaedic Research Center, UCLA & Orthopaedic Hospital Department of Orthopaedic Surgery, 2400 S Flower Street, Los Angeles, CA 90007, USA.
Clin Orthop Relat Res. 2011 Feb;469(2):395-404. doi: 10.1007/s11999-010-1555-6.
The low wear rates of crosslinked polyethylenes provide the potential to use larger diameters to resist dislocation. However, this requires the use of thinner liners in the acetabular component, with concern that higher contact stresses will increase wear, offsetting the benefits of the crosslinking.
QUESTIONS/PURPOSES: We asked the following questions: Is the wear of conventional and crosslinked polyethylene liners affected by ball diameter, rigidity of backing, and liner thickness? Are the stresses in the liner affected by thickness?
Wear rates were measured in a hip simulator and stresses were calculated using finite element modeling.
Without crosslinking, the wear rate was 4% to 10% greater with a 36-mm diameter than a 28-mm diameter. With crosslinking, wear was 9% lower with a 36-mm diameter without metal backing and 4% greater with metal backing. Reducing the thickness from 6 mm to 3 mm increased the contact stress by 46%, but the wear rate decreased by 19%.
The reduction in wear with 5 Mrad of crosslinking was not offset by increasing the diameter from 28 mm to 36 mm or by using a liner as thin as 3 mm.
The results indicate, for a properly positioned 5-Mrad crosslinked acetabular component and within the range of dimensions evaluated, neither wear nor stresses in the polyethylene are limiting factors in the use of larger-diameter, thinner cups to resist dislocation.
交联聚乙烯的低磨损率为使用更大直径来抵抗脱位提供了潜力。然而,这需要在髋臼部件中使用更薄的衬垫,因为担心更高的接触应力会增加磨损,从而抵消交联的益处。
问题/目的:我们提出了以下问题:球直径、衬垫背面的刚性和衬垫厚度是否会影响传统和交联聚乙烯衬垫的磨损?衬垫的厚度是否会影响衬垫中的应力?
在髋关节模拟器中测量磨损率,并使用有限元建模计算应力。
未交联时,36 毫米直径的磨损率比 28 毫米直径的磨损率高 4%至 10%。交联后,无金属背衬的 36 毫米直径的磨损率降低了 9%,而有金属背衬的磨损率增加了 4%。衬垫厚度从 6 毫米减至 3 毫米会使接触应力增加 46%,但磨损率降低 19%。
5 Mrad 的交联虽降低了磨损,但增加直径(从 28 毫米增至 36 毫米)或使用厚度仅为 3 毫米的衬垫并未抵消这种效果。
结果表明,对于正确定位的 5 Mrad 交联髋臼部件,并且在评估的尺寸范围内,使用更大直径、更薄的髋臼杯来抵抗脱位,聚乙烯中的磨损和应力都不是限制因素。