Validity and limits of the rebound tonometer (ICare®): clinical study.

作者信息

Scuderi Gian Luca, Cascone Nikhil Carlo, Regine Federico, Perdicchi Andrea, Cerulli Angelica, Recupero Santi Maria

机构信息

Department of Ophthalmology, University of Rome La Sapienza 2nd School of Medicine, Sant'Andrea Hospital, Rome, Italy.

出版信息

Eur J Ophthalmol. 2011 May-Jun;21(3):251-7. doi: 10.5301/EJO.2010.3712.

Abstract

PURPOSE

The aim of this study was to evaluate the measurement of intraocular pressure (IOP) using a new induction/impact rebound tonometer (ICare®) compared with Goldmann applanation tonometry (GAT). We also aimed to quantify the systematic and random errors (bias) of the 2 methods, to evaluate the sensitivity and specificity of the ICare® tonometer in identifying patients with 21 mmHg or more measured with the GAT, and to study the influence of corneal thickness on IOP measurement with the 2 tonometers.

METHODS

We compared the IOP values obtained with the 2 instruments in 97 patients.

RESULTS

Analysis based on the Bland and Altman method revealed that the IOP values recorded with the ICare® tonometer were slightly higher than those obtained with the GAT. The estimated bias for right eye measurements was 0.78 mmHg with 95% limits of agreement ±3.55 mmHg. This overestimation, which is not clinically relevant, was confirmed when we used the IOP values corrected according to central corneal thickness for data analysis. The sensitivity and specificity were 0.90 and 0.95, respectively.

CONCLUSIONS

The ICare® tonometer proved to be comparable with other nonconventional tonometers and can be used by nonophthalmologists and paramedical personnel during screening tests of populations. In addition, the ICare® tonometer could be considered a valid alternative to GAT when GAT is not available.

摘要

目的

本研究旨在评估一种新型感应/冲击回弹眼压计(ICare®)与Goldmann压平眼压计(GAT)测量眼压(IOP)的情况。我们还旨在量化这两种方法的系统误差和随机误差(偏差),评估ICare®眼压计在识别GAT测量值为21 mmHg或更高的患者时的敏感性和特异性,并研究角膜厚度对这两种眼压计测量IOP的影响。

方法

我们比较了97例患者使用这两种仪器获得的IOP值。

结果

基于Bland和Altman方法的分析显示,ICare®眼压计记录的IOP值略高于GAT获得的值。右眼测量的估计偏差为0.78 mmHg,95%一致性界限为±3.55 mmHg。当我们使用根据中央角膜厚度校正的IOP值进行数据分析时,证实了这种临床上无关紧要的高估。敏感性和特异性分别为0.90和0.95。

结论

ICare®眼压计被证明与其他非传统眼压计相当,可由非眼科医生和辅助医疗人员在人群筛查测试中使用。此外,当无法使用GAT时,ICare®眼压计可被视为GAT的有效替代方法。