家庭自眼压测量(iCare HOME(TA022))在健康人群和青光眼患者眼压测量中的性能。

Performance of home-based self-tonometry (iCare HOME (TA022)) for measuring intraocular pressure among healthy and glaucoma patients.

机构信息

Department of Allied Health Sciences,Faculty of Life & Allied Health Sciences(FLAHS), Ramaiah University of Applied Sciences, Bangalore, Karnataka, 560054, India.

Department of Optometry, Manipal College of Health Professions (MCHP), Manipal Academy of Higher Education (MAHE), Manipal, Udupi (Dist), Karnataka, 576104, India.

出版信息

F1000Res. 2023 Sep 22;12:128. doi: 10.12688/f1000research.123104.2. eCollection 2023.

Abstract

The purpose of this study was to compare iCare HOME (TA022) with Goldmann applanation tonometer and to evaluate the self-tonometry measurements among the Indian population. Eligible patients underwent iCare HOME training through guided demonstration (verbal, pictorial, video) and practised self-tonometry measures using iCare HOME. Certification for independent iCare HOME measure was provided if first iCare HOME intraocular pressure (IOP) measurement fell within ± 5 mmHg of Goldmann applanation tonometer (GAT) measurement which was measured by the trained clinician (principal investigator). Certified participants underwent simulated home self-tonometry measurements using iCare HOME, and agreement with GAT IOP measurements was assessed. Seven of 83 participants (8.43%) failed to complete the study due to difficulty in performing the task, leading to non-certification. Patients who could use the iCare HOME had a mean age of 53 ± 15.55years (53% males; 46% females). Only one in 12 subjects did not qualify to use iCare HOME. The overall mean difference between iCare HOME and GAT was 0.83 mmHg (95%, 3.92 and -2.25). At various pressure ranges, 7-16 mmHg, 17-23 mmHg and >23 mmHg, the mean difference between iCare HOME and GAT was 1.22 mmHg (95%, 4.32 and -1.86), 0.77 mmHg (95%,3.69 and -2.19), -0.11 mmHg (95%, 2.52 and -2.74) respectively. The intra-class correlation coefficient of the iCare HOME device was 0.997(95% CI,0.995-0.998). Patients were able to perform self- tonometry using iCare HOME with good reliability and safety. iCare HOME can be used to address the issue of difficulty in acquiring frequent and diurnal IOP measurements by patients doing self-tonometry from home.

摘要

本研究旨在比较 iCare HOME(TA022)与 Goldmann 压平眼压计,并评估其在印度人群中的自我眼压测量值。符合条件的患者通过指导演示(口头、图片、视频)接受 iCare HOME 培训,并使用 iCare HOME 进行自我眼压测量练习。如果首次 iCare HOME 眼压(IOP)测量值与经过训练的临床医生(主要研究者)测量的 Goldmann 压平眼压计(GAT)测量值相差±5mmHg,则提供独立 iCare HOME 测量认证。经过认证的参与者使用 iCare HOME 进行模拟家庭自我眼压测量,并评估与 GAT IOP 测量值的一致性。由于难以完成任务,有 7 名(8.43%)患者未能完成研究,导致未获得认证。能够使用 iCare HOME 的患者平均年龄为 53±15.55 岁(53%为男性;46%为女性)。仅有 12 名患者中的 1 名不符合使用 iCare HOME 的条件。iCare HOME 与 GAT 的总体平均差异为 0.83mmHg(95%,3.92mmHg 和-2.25mmHg)。在不同的压力范围内,7-16mmHg、17-23mmHg 和>23mmHg,iCare HOME 与 GAT 的平均差值分别为 1.22mmHg(95%,4.32mmHg 和-1.86mmHg)、0.77mmHg(95%,3.69mmHg 和-2.19mmHg)、-0.11mmHg(95%,2.52mmHg 和-2.74mmHg)。iCare HOME 设备的组内相关系数为 0.997(95%置信区间,0.995-0.998)。患者能够使用 iCare HOME 进行自我眼压测量,具有良好的可靠性和安全性。iCare HOME 可用于解决患者在家中进行自我眼压测量时难以频繁和日间获取眼压测量值的问题。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/dfe7/11184362/ac7a297f1ccc/f1000research-12-156214-g0000.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验