Louis Stokes Veterans Affairs Medical Center, Cleveland, OH 44109, USA.
IEEE Trans Neural Syst Rehabil Eng. 2011 Feb;19(1):45-53. doi: 10.1109/TNSRE.2010.2079952. Epub 2010 Sep 27.
An implantable stimulator-telemeter (IST-12) was developed for applications in neuroprosthetic restoration of limb function in paralyzed individuals. The IST-12 provides 12 stimulation channels and two myoelectric signal (MES) channels. The MES circuitry includes a two-channel multiplexer, preamplifier, variable gain amplifier/bandpass filter, full-wave rectifier, and bin integrator. Power and control signals are transmitted from an external control unit to the IST-12 through an inductive link. Recorded MES signals are telemetered back to the external control unit through the same inductive link. Following bench testing, one device was implanted chronically in a dog for 15 months and evaluated. Conditions were identified in which MES could be recorded with minimal stimulus artifact. The ability to record MES in the presence of stimulation was verified, confirming the potential of the IST-12 to be used as a myoelectric controlled neuroprosthesis.
一种可植入的刺激遥测器(IST-12)被开发出来,用于在瘫痪患者中恢复肢体功能的神经假肢。IST-12 提供 12 个刺激通道和两个肌电信号(MES)通道。MES 电路包括双通道复用器、前置放大器、可变增益放大器/带通滤波器、全波整流器和二进制积分器。电源和控制信号通过感应链路从外部控制单元传输到 IST-12。记录的 MES 信号通过相同的感应链路遥测回外部控制单元。经过台架测试,一个设备在一只狗体内被植入并进行了 15 个月的慢性评估。确定了在最小刺激伪影的情况下可以记录 MES 的条件。验证了在刺激存在的情况下记录 MES 的能力,这证实了 IST-12 作为肌电控制神经假肢的潜力。