Halpern E J, Esser P D, Nickoloff E L, Alderson P O
Department of Radiology, Columbia Presbyterian Medical Center, New York, NY 10032.
J Digit Imaging. 1990 Feb;3(1):42-8. doi: 10.1007/BF03168110.
Quality control is fundamental to the clinical application of digital radiography. A 14 x 17-in phantom radiograph was designed to test digital image quality by measurement of five parameters: high-contrast spatial resolution, low-contrast discrimination, linearity of gray-scale response, high-frequency noise, and geometric distortion. The phantom was used to evaluate the AT&T-Philips CommView picture archival and communications system (AT&T Bell Laboratories, West Long Branch, NJ; Philips Medical Systems, Shelton, CT). High-contrast resolution was found to be greater along the diagonal axis of the system than along either the horizontal or vertical axis. Problems with low-contrast discrimination and linearity of gray-scale response were identified. This phantom provides a simple tool for daily quality assurance testing and an objective standard for comparison of image quality between different digital radiography systems.