Palenik C J, Adams M L, Miller C H
Department of Oral Microbiology, Indiana University, Indianapolis.
Am J Dent. 1990 Dec;3(6):239-44.
The effectiveness of steam autoclaving on bacterial endospores placed within five types of sharps containers was tested. A variety of container physical orientations within the autoclave were evaluated. Spores were present on commercial spore strips or placed onto capped and uncapped dental needles. All strips and needles present in empty or filled containers could be sterilized within 15 minutes when the containers were placed on their sides and their vents left open. The contents of containers processed in an upward position required between 30-60 minutes of autoclaving before being sterilized. The size and shape of the containers influenced ease of sterilization.
测试了蒸汽高压灭菌法对放置在五种锐器容器内的细菌芽孢的有效性。评估了高压灭菌器内各种容器的物理方向。芽孢存在于商用芽孢条上,或放置在有帽和无帽的牙科针上。当容器侧放且通风口打开时,空容器或装满物品的容器内所有的条带和针在15分钟内均可被灭菌。以向上位置处理的容器内容物在灭菌前需要30 - 60分钟的高压灭菌时间。容器的尺寸和形状影响灭菌的难易程度。