Effectiveness of three types of sterilization on the contents of sharps containers.

作者信息

Sheldrake M A, Majors C D, Gaines D J, Palenik C J

机构信息

Infection Control Research and Services, Indiana University, School of Dentistry, Indianapolis 46202, USA.

出版信息

Quintessence Int. 1995 Nov;26(11):771-8.

PMID:8628836
Abstract

The purpose of this study was to test the effect of treatment in a gravity steam autoclave, high-vacuum steam autoclave, or an unsaturated chemical vapor sterilizer on endospores present on strips or placed inside of dental anesthetic cartridges held within sharps containers. Strips with 1.7 X 10(5) Bacillus stearothermophilus endospores were used; the cartridges were soiled with an equal number of spores or spores mixed with blood. If sterilization was not accomplished after the initial period, additional exposure time was added. Neither the presence of blood or anesthetic solution nor the position of the container affected the efficiency of sterilization. Soiled cartridges were much more difficult than strips to sterilize. Intact cartridges could not be sterilized by two runs in a gravity steam autoclave or an unsaturated chemical vapor sterilizer or one run in a high-vacuum steam autoclave. Sterilization occurred after two runs in the gravity steam autoclave and unsaturated chemical vapor sterilizer only when one end of the cartridge was removed prior to processing. Results indicated that unopened spore-soiled cartridges are not readily sterilized by commonplace office sterilizers, even after extended exposure.

摘要

本研究的目的是测试重力蒸汽灭菌器、高真空蒸汽灭菌器或不饱和化学蒸汽灭菌器对存在于条带上或放置在锐器容器内牙科麻醉药筒内部的芽孢的处理效果。使用带有1.7×10⁵嗜热脂肪芽孢杆菌芽孢的条带;药筒被等量的芽孢或与血液混合的芽孢弄脏。如果在初始阶段后未完成灭菌,则增加额外的暴露时间。血液或麻醉溶液的存在以及容器的位置均不影响灭菌效率。弄脏的药筒比条带更难灭菌。完整的药筒在重力蒸汽灭菌器或不饱和化学蒸汽灭菌器中进行两次运行,或在高真空蒸汽灭菌器中进行一次运行都无法灭菌。仅当药筒的一端在处理前被移除时,在重力蒸汽灭菌器和不饱和化学蒸汽灭菌器中进行两次运行后才会发生灭菌。结果表明,即使经过长时间暴露,未开封的被芽孢污染的药筒也不容易被普通的办公室灭菌器灭菌。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验