Department of Medical Microbiology and Infectious Diseases, Canisius Wilhelmina Hospital, and Radboud University Nijmegen Medical Center, Weg door Jonkerbos 100, 6532 SZ Nijmegen, Netherlands.
J Clin Microbiol. 2010 Dec;48(12):4634-5. doi: 10.1128/JCM.00464-10. Epub 2010 Oct 20.
Contamination of an in-house diagnostic real-time PCR for Q fever was traced back to a commercially obtained PCR Master Mix. It was established that this Master Mix contained DNA from Coxiella burnetii, probably as a result of the use of compounds of animal origin such as bovine serum albumin.
室内诊断实时 PCR 检测 Q 热的污染可追溯到市售的 PCR 主混合液。据证实,这种主混合液中含有贝氏柯克斯体的 DNA,可能是由于使用了牛血清白蛋白等动物来源的化合物。