Department of Radiation Oncology, M. D. Anderson Cancer Center Orlando, Orlando, Florida 32806, USA.
Med Phys. 2010 Sep;37(9):4817-53. doi: 10.1118/1.3462971.
Helical tomotherapy is a relatively new modality with integrated treatment planning and delivery hardware for radiation therapy treatments. In view of the uniqueness of the hardware design of the helical tomotherapy unit and its implications in routine quality assurance, the Therapy Physics Committee of the American Association of Physicists in Medicine commissioned Task Group 148 to review this modality and make recommendations for quality assurance related methodologies. The specific objectives of this Task Group are: (a) To discuss quality assurance techniques, frequencies, and tolerances and (b) discuss dosimetric verification techniques applicable to this unit. This report summarizes the findings of the Task Group and aims to provide the practicing clinical medical physicist with the insight into the technology that is necessary to establish an independent and comprehensive quality assurance program for a helical tomotherapy unit. The emphasis of the report is to describe the rationale for the proposed QA program and to provide example tests that can be performed, drawing from the collective experience of the task group members and the published literature. It is expected that as technology continues to evolve, so will the test procedures that may be used in the future to perform comprehensive quality assurance for helical tomotherapy units.
螺旋断层放疗是一种具有集成治疗计划和治疗硬件的相对较新的放射治疗模式。鉴于螺旋断层放疗设备硬件设计的独特性及其对常规质量保证的影响,美国医学物理学家协会治疗物理委员会委托第 148 任务组审查该模式,并为质量保证相关方法提出建议。该任务组的具体目标是:(a)讨论质量保证技术、频率和容差,以及 (b)讨论适用于该设备的剂量验证技术。本报告总结了任务组的调查结果,旨在为临床实践中的医学物理学家提供必要的技术洞察力,以建立独立且全面的螺旋断层放疗设备质量保证计划。报告的重点是描述拟议的 QA 计划的基本原理,并提供可以执行的示例测试,这些测试是从任务组成员的集体经验和已发表的文献中得出的。预计随着技术的不断发展,未来可能会使用哪些测试程序来对螺旋断层放疗设备进行全面的质量保证。