J Am Coll Radiol. 2010 Nov;7(11):878-84. doi: 10.1016/j.jacr.2010.06.019.
MRI is increasingly used for the detection of breast carcinoma. Little is known about breast MRI techniques among community practice facilities. The aim of this study was to evaluate equipment and acquisition techniques used by community facilities across the United States, including compliance with minimum standards by the ACRIN® 6667 Trial and the European Society of Breast Imaging.
Breast Cancer Surveillance Consortium facilities performing breast MRI were identified and queried by survey regarding breast MRI equipment and technical parameters. Variables included scanner field strength, coil type, acquisition coverage, slice thickness, and the timing of the initial postcontrast sequence. Results were tallied and percentages of facilities meeting ACRIN® and European Society of Breast Imaging standards were calculated.
From 23 facilities performing breast MRI, results were obtained from 14 (61%) facilities with 16 MRI scanners reporting 18 imaging parameters. Compliance with equipment recommendations of ≥1.5-T field strength was 94% and of a dedicated breast coil was 100%. Eighty-three percent of acquisitions used bilateral postcontrast techniques, and 78% used slice thickness≤3 mm. The timing of initial postcontrast sequences ranged from 58 seconds to 8 minutes 30 seconds, with 63% meeting recommendations for completion within 4 minutes.
Nearly all surveyed facilities met ACRIN and European Society of Breast Imaging standards for breast MRI equipment. The majority met standards for acquisition parameters, although techniques varied, in particular for the timing of initial postcontrast imaging. Further guidelines by the ACR Breast MRI Accreditation Program will be of importance in facilitating standardized and high-quality breast MRI.
MRI 越来越多地用于乳腺癌的检测。但社区实践设施对乳腺 MRI 技术知之甚少。本研究的目的是评估美国各地社区设施使用的设备和采集技术,包括符合 ACRIN®6667 试验和欧洲乳腺成像协会的最低标准。
通过调查,确定了进行乳腺 MRI 的乳腺癌监测联盟设施,并询问了有关乳腺 MRI 设备和技术参数的问题。变量包括扫描仪场强、线圈类型、采集范围、切片厚度以及初始对比后序列的时间。对结果进行了统计,并计算了符合 ACRIN®和欧洲乳腺成像协会标准的设施的百分比。
在进行乳腺 MRI 的 23 个设施中,有 14 个(61%)设施的结果来自 16 个 MRI 扫描仪,报告了 18 个成像参数。≥1.5-T 场强设备建议的符合率为 94%,专用乳腺线圈的符合率为 100%。83%的采集采用双侧对比后技术,78%采用≤3mm 的切片厚度。初始对比后序列的时间范围从 58 秒到 8 分 30 秒,63%的时间符合 4 分钟内完成的建议。
几乎所有接受调查的设施都符合 ACRIN 和欧洲乳腺成像协会的乳腺 MRI 设备标准。大多数设施都符合采集参数标准,但技术各不相同,特别是在初始对比后成像的时间方面。ACR 乳腺 MRI 认证计划的进一步指南将对促进标准化和高质量的乳腺 MRI 具有重要意义。