CIC-P INSERM 803, Plateforme d'Investigation Technologique, Dijon University Hospital, Dijon F-21079, France.
Osteoarthritis Cartilage. 2011 Feb;19(2):194-9. doi: 10.1016/j.joca.2010.10.024. Epub 2010 Nov 5.
(1) to investigate the test-retest reliability of 3D gait analysis (3DGA) in hip Osteoarthritis (OA) patients; (2) to find the minimum number of gait trials needed to overcome intrinsic variability; (3) to check the accuracy of angles measured by the 3D system.
23 Patients suffering from hip OA with no other major disease were recruited. We evaluated the reliability of spatio-temporal variables and body angles (lower-limb joints, trunk and pelvis angles) during two sessions of 3DGA using intra-class correlation coefficients (ICC). The minimum number of trials needed to overcome intrinsic variability was evaluated using an exponential fit model and the Bland and Altman coefficient of repeatability (CoR). The accuracy of measurement was evaluated using a manual goniometer and the recording of 18 different angles.
Spatio-temporal variables and most of the kinematic joint and trunk angles calculated demonstrated good to excellent reliability (ICC from 0.77 to 0.97). This was not the case for pelvic angles. The fitting model combined with the CoR showed that 5-10 trials are sufficient to obtain good reliability [ICC>0.7; CoR<2 standard deviation (SD)] for most of the spatio-temporal variables. All body angles showed good reliability (ICC>0.7) and low CoR (<2 SD) after five trials except for the pelvic angles. The reliability of marker positioning was found to be good (ICC>0.7) to excellent (ICC>0.9). Differences between angles measured using 3DGA and angles measured with a manual goniometer were found to be less than one percent.
The present study shows that most of variables obtained using 3DGA in hip OA patients are reliable. Moreover, for most variables, 5-10 trials are needed to obtain good reliability and to overcome intrinsic variability, rather than 30 or more, thus improving the feasibility of measurement.
(1)研究 3D 步态分析(3DGA)在髋骨关节炎(OA)患者中的测试-重测可靠性;(2)找出克服内在变异性所需的最小步态试验次数;(3)检查 3D 系统测量角度的准确性。
招募了 23 名患有髋骨 OA 且无其他重大疾病的患者。我们使用组内相关系数(ICC)评估了两次 3DGA 期间时空变量和身体角度(下肢关节、躯干和骨盆角度)的可靠性。使用指数拟合模型和 Bland 和 Altman 重复性系数(CoR)评估了克服内在变异性所需的最小试验次数。使用手动量角器和记录 18 个不同角度评估了测量的准确性。
时空变量和大多数计算出的运动学关节和躯干角度显示出良好到极好的可靠性(ICC 从 0.77 到 0.97)。骨盆角度则不然。拟合模型与 CoR 相结合表明,大多数时空变量需要 5-10 次试验才能获得良好的可靠性[ICC>0.7;CoR<2 标准差(SD)]。除了骨盆角度外,所有身体角度在进行五次试验后均显示出良好的可靠性(ICC>0.7)和较低的 CoR(<2 SD)。标记定位的可靠性被发现是良好的(ICC>0.7)到极好的(ICC>0.9)。使用 3DGA 测量的角度与使用手动量角器测量的角度之间的差异小于 1%。
本研究表明,髋骨 OA 患者使用 3DGA 获得的大多数变量都是可靠的。此外,对于大多数变量,需要 5-10 次试验才能获得良好的可靠性并克服内在变异性,而不是 30 次或更多,从而提高了测量的可行性。