Department of Radiology, The Children's Hospital of Philadelphia, 34th Street and Civic Center Boulevard, Philadelphia, PA 19104, USA.
Pediatr Radiol. 2011 Apr;41(4):476-82. doi: 10.1007/s00247-010-1885-2. Epub 2010 Nov 16.
Manufacturers have provided C-arm CT imaging technologies for applications in interventional radiology in recent years. However, clinical imaging protocols and radiation doses have not been well studied or reported.
The purpose of this study is to develop low-dose settings for clinically acceptable CT imaging of temporomandibular joint in interventional radiology suites, using a C-arm imaging angiography system.
CT scans were performed with a flat-panel digital C-arm angiographic system on a 5-year-old anthropomorphic phantom. The CTDI was determined for various rotation times, dose settings and Cu filter selections. The CTDI values were compared with those of conventional low-dose CT for the same phantom. The effectiveness of using Cu filters to reduce dose was also investigated. Images were reviewed by a senior radiologist for clinical acceptance.
The manufacturer's default setting gave an equivalent CTDI of 4.8 mGy. Optimizing the dose settings and adding copper filtration reduced the radiation dose by 94%. This represents a 50% reduction from conventional CT.
Use of Cu filters and low-dose settings significantly reduced radiation dose from that of standard settings. This phantom study process successfully guided the clinical implementation of low-dose studies for all ages at our institution.
近年来,制造商为介入放射学应用提供了 C 臂 CT 成像技术。然而,临床成像协议和辐射剂量尚未得到很好的研究或报道。
本研究旨在使用 C 臂成像血管造影系统为介入放射学套房中的颞下颌关节开发临床可接受的 CT 成像的低剂量设置。
在一个 5 岁的人体模型上使用平板数字 C 臂血管造影系统进行 CT 扫描。确定了各种旋转时间、剂量设置和 Cu 滤光片选择的 CTDI。将 CTDI 值与相同的模型的常规低剂量 CT 进行比较。还研究了使用 Cu 滤光片降低剂量的效果。由一位资深放射科医生对图像进行了临床评估。
制造商的默认设置给出了等效 CTDI 为 4.8 mGy。优化剂量设置并添加铜过滤可将辐射剂量降低 94%。这代表与常规 CT 相比降低了 50%。
使用 Cu 滤光片和低剂量设置可显著降低标准设置的辐射剂量。这项体模研究过程成功地指导了我们机构中所有年龄段的低剂量研究的临床实施。