Department of Radiation Oncology, University of Michigan, Ann Arbor, USA.
J Appl Clin Med Phys. 2010 Aug 5;11(4):3172. doi: 10.1120/jacmp.v11i4.3172.
Film calibration is time-consuming work when dose accuracy is essential while working in a range of photon scatter environments. This study uses the single-target single-hit model of film response to fit the calibration curves as a function of calibration method, processor condition, field size and depth. Kodak XV film was irradiated perpendicular to the beam axis in a solid water phantom. Standard calibration films (one dose point per film) were irradiated at 90 cm source-to-surface distance (SSD) for various doses (16-128 cGy), depths (0.2, 0.5, 1.5, 5, 10 cm) and field sizes (5 × 5, 10 × 10 and 20 × 20 cm²). The 8-field calibration method (eight dose points per film) was used as a reference for each experiment, taken at 95 cm SSD and 5 cm depth. The delivered doses were measured using an Attix parallel plate chamber for improved accuracy of dose estimation in the buildup region. Three fitting methods with one to three dose points per calibration curve were investigated for the field sizes of 5 × 5, 10 × 10 and 20 × 20 cm². The inter-day variation of model parameters (background, saturation and slope) were 1.8%, 5.7%, and 7.7% (1 σ) using the 8-field method. The saturation parameter ratio of standard to 8-field curves was 1.083 ± 0.005. The slope parameter ratio of standard to 8-field curves ranged from 0.99 to 1.05, depending on field size and depth. The slope parameter ratio decreases with increasing depth below 0.5 cm for the three field sizes. It increases with increasing depths above 0.5 cm. A calibration curve with one to three dose points fitted with the model is possible with 2% accuracy in film dosimetry for various irradiation conditions. The proposed fitting methods may reduce workload while providing energy dependence correction in radiographic film dosimetry. This study is limited to radiographic XV film with a Lumisys scanner.
胶片校准是一项耗时的工作,尤其是在各种光子散射环境中需要精确剂量时。本研究使用胶片响应的单靶单点模型来拟合校准曲线,作为校准方法、处理器条件、射野大小和深度的函数。柯达 XV 胶片垂直于束轴在固体水模体中照射。标准校准胶片(每张胶片一个剂量点)在 90cm 源皮距(SSD)处照射不同剂量(16-128cGy)、深度(0.2、0.5、1.5、5、10cm)和射野大小(5×5、10×10 和 20×20cm²)。8 野校准方法(每张胶片 8 个剂量点)作为每个实验的参考,在 95cm SSD 和 5cm 深度处进行。使用 Attix 平行板电离室测量传递剂量,以提高在建成区剂量估计的准确性。研究了三种拟合方法,每种方法的校准曲线有一个至三个剂量点,用于 5×5、10×10 和 20×20cm² 的射野大小。使用 8 野方法,模型参数(背景、饱和和斜率)的日间变化为 1.8%、5.7%和 7.7%(1σ)。标准和 8 野曲线的饱和参数比为 1.083±0.005。标准和 8 野曲线的斜率参数比在 0.99 至 1.05 之间,取决于射野大小和深度。对于三个射野大小,在 0.5cm 以下的深度,斜率参数比随深度增加而减小;在 0.5cm 以上的深度,斜率参数比随深度增加而增加。对于各种照射条件,使用模型拟合一个至三个剂量点的校准曲线,可以达到 2%的胶片剂量准确性。在放射胶片剂量学中,所提出的拟合方法可以减少工作量,同时提供能量依赖性校正。本研究仅限于使用 Lumisys 扫描仪的放射型 XV 胶片。