一种基于仪器分析咬合及其在牙弓中产生的力分布方法的精度。

Precision of an instrumentation-based method of analyzing occlusion and its resulting distribution of forces in the dental arch.

作者信息

Koos Bernd, Godt Arnim, Schille Christine, Göz Gernot

机构信息

Department of Orthodontics, University Hospital and Medical Faculty of Eberhard Karl University, Tübingen, Germany.

出版信息

J Orofac Orthop. 2010 Nov;71(6):403-10. doi: 10.1007/s00056-010-1023-7. Epub 2010 Nov 17.

Abstract

BACKGROUND AND OBJECTIVE

In practice, analysis of occlusion is reduced to depicting it with color-marking foils. Precise analysis that incorporates time resolution and plots the distribution of forces within the occlusion is not possible in the everyday clinical situation with the usual methods. T-Scan(®) III is a computer-assisted dental occlusion analyzer that depicts occlusion by means of pressure-sensitive foils. The aim of our study was to test the accuracy and reliability of this method.

SUBJECTS AND METHODS

The study population comprised 42 subjects (23 male and 19 female, aged 20-30, median age 26 years). The measurements were performed using the TScan(®) III from Tekscan Inc., South Boston, MA, USA. Six recordings with two foils were made for each subject and a total of 30 masticatory cycles were registered. Statistical analysis referred to the method's measurement accuracy and reliability, as well as the influence of changing the foil and repositioning the T-Scan(®) III during the repeated measurements.

RESULTS

The percentage distribution of forces per tooth ranged from 0 to 41%. The mean measurement per tooth was 6.9% of the maximum total force exerted. The measurement error was 1%, the 1.96-fold measurement error calculated according to Bland & Altman (accuracy) was 2% and the 2.77-fold measurement error (reliability) was 2.8%. Neither changing the foil nor the repeated measuring had any statistically significant influences on the measured value.

CONCLUSION

The measuring technique studied is superior to the usual methods, particularly with regard to force analysis per tooth. The level of accuracy is acceptable and no interference arising from change of foil or repeated measuring was detected. The method presented in this study therefore enhances routine diagnostics with marking foils. A combination of this method with marking foils would be ideal because the pressure-sensitive foils in this system do not produce any contact markings intraorally. This combination enables the contacts depicted on the computer to be assigned intraorally with even greater precision.

摘要

背景与目的

在实际操作中,咬合分析简化为使用彩色标记箔来描绘。在日常临床情况下,采用常规方法无法进行结合时间分辨率并绘制咬合内力量分布的精确分析。T-Scan(®) III是一种计算机辅助的牙齿咬合分析仪,通过压敏箔描绘咬合情况。我们研究的目的是测试该方法的准确性和可靠性。

对象与方法

研究对象包括42名受试者(23名男性和19名女性,年龄20 - 30岁,中位年龄26岁)。使用美国马萨诸塞州南波士顿Tekscan公司的T-Scan(®) III进行测量。为每位受试者使用两张箔进行6次记录,共记录30个咀嚼周期。统计分析涉及该方法的测量准确性和可靠性,以及在重复测量过程中更换箔和重新定位T-Scan(®) III的影响。

结果

每颗牙齿的力量百分比分布范围为0至41%。每颗牙齿的平均测量值为施加的最大总力量的6.9%。测量误差为1%,根据布兰德和奥特曼方法计算的1.96倍测量误差(准确性)为2%,2.77倍测量误差(可靠性)为2.8%。更换箔或重复测量对测量值均无统计学上的显著影响。

结论

所研究的测量技术优于常规方法,特别是在每颗牙齿的力量分析方面。准确性水平是可接受的,未检测到因更换箔或重复测量而产生的干扰。因此,本研究中提出的方法增强了使用标记箔的常规诊断。将该方法与标记箔相结合将是理想的,因为该系统中的压敏箔在口腔内不会产生任何接触标记。这种结合能够更精确地在口腔内确定计算机上描绘的接触点。