Friedrich David, Aubert Xavier L, Fuhr Hartmut, Brauers Andreas
Philips Research Europe, Lehrstuhl für Bildverarbeitung, Sommerfeldstr. 24, 52074 Aachen, Germany.
Annu Int Conf IEEE Eng Med Biol Soc. 2010;2010:4048-51. doi: 10.1109/IEMBS.2010.5627626.
We present an algorithm for obtaining the heart rate from the signal of a single, contact-less sensor recording the mechanical activity of the heart. This vital parameter is required on a beat-to-beat basis for applications in sleep analysis and heart failure disease management. Our approach bundles information from various sources for first robust estimates. These estimates are further refined in a second step. An unambiguous comparison with the ECG RR-intervals taken as reference is possible for 98.5% of the heart beats. In these cases, a mean absolute error of 17 ms for the inter-beat interval lengths has been achieved, over a test corpus of 20 whole nights.