Hansen J E, Casaburi R, Crapo R O, Jensen R L
Department of Medicine, UCLA School of Medicine, Torrance.
Am Rev Respir Dis. 1990 May;141(5 Pt 1):1190-3. doi: 10.1164/ajrccm/141.5_Pt_1.1190.
Blood gas proficiency testing has focused on assessing the accuracy of measurement of each analyte (pH, PCO2) independently of each other. Recently, the American Thoracic Society-California Thoracic Society Blood Gas Proficiency Testing Survey distributed the same lot of ampules of proficiency testing material (a buffered fluorocarbon-containing emulsion) on three occasions within a 1-yr period, allowing us to assess the precision (reproducibility) of measurement of each analyte. Comparing 580 instruments of 13 models, we found that the precision of measurement of each analyte was positively correlated with the precision of measurement of each other analyte, and the correlation of precision between models was much stronger than precision between the individual instruments. We also found correlation of precision of each analyte with two targets for accuracy: (1) the all-instrument mean and (2) the model-specific means. Correlations were higher with the model-specific means. These findings suggest: (1) that features unique to design of each model are important in the precision of measurement of these ampules, and (2) that it would be informative to include measurements of precision with linked and cumulative ratings of analyte accuracy in proficiency testing rating systems.
血气能力验证一直专注于独立评估每种分析物(pH、PCO2)测量的准确性。最近,美国胸科学会 - 加利福尼亚胸科学会血气能力验证测试调查在1年时间内分三次分发了同一批次的能力验证测试材料安瓿(一种含缓冲剂的含氟碳乳液),这使我们能够评估每种分析物测量的精密度(再现性)。通过比较13种型号的580台仪器,我们发现每种分析物测量的精密度与其他每种分析物测量的精密度呈正相关,并且不同型号之间精密度的相关性比单个仪器之间的精密度相关性要强得多。我们还发现每种分析物的精密度与两个准确性目标相关:(1)所有仪器的平均值和(2)特定型号的平均值。与特定型号的平均值相关性更高。这些发现表明:(1)每种型号设计的独特特征对于这些安瓿测量的精密度很重要,(2)在能力验证测试评级系统中纳入精密度测量以及分析物准确性的关联和累积评级将有助于提供信息。