Itano M
Am J Clin Pathol. 1983 Oct;80(4 Suppl):554-62.
The results of the 1982 CAP Blood Gas Survey, in which fluorocarbon-based emulsion specimens were used, are compared to the 1981 results, in which aqueous specimens were used. Interinstrument differences exist with all three analytes, pH PCO2, and PO2. There was little difference in interlaboratory precision of pH and PCO2 determinations in the two surveys. The coefficients of variation for PO2 determination were lower in 1982 than in 1981. This is attributed to the oxygen buffering capacity of fluorocarbon. Without a reference method for measuring blood pH, PCO2, and PO2, precision and interinstrument differences can be measured, but accuracy cannot.
1982年临床检验标准化委员会(CAP)血气调查使用了基于氟碳化合物的乳剂标本,其结果与1981年使用水性标本的结果进行了比较。三种分析物(pH、PCO₂和PO₂)在不同仪器之间均存在差异。两次调查中pH和PCO₂测定的实验室间精密度差异不大。1982年PO₂测定的变异系数低于1981年。这归因于氟碳化合物的氧缓冲能力。在没有测量血液pH、PCO₂和PO₂的参考方法的情况下,可以测量精密度和仪器间差异,但无法测量准确性。