Aseptic loosening of cemented hip prostheses is recognized as a long-term problem, and especially in males and younger patients. Much energy has been focused on developing new prostheses that are designed for cementless fixation. We evaluated the performance of and periprosthetic bone response to a tapered, titanium, hydroxyapatite (HA)-coated femoral hip prosthesis at a minimum of 7 years of follow-up after treatment with primary total hip arthroplasty.
Seventy-eight patients and 86 hips were included in the study. There were 35 men and 43 women; the mean age at the time of the operation was 59 years (range, 41 to 81 years). We used a tapered, titanium (Ti6Al4V), HA-coated femoral implant. We evaluated the patients at a minimum of 7 years of follow-up after treatment with primary total hip arthroplasty. Clinical evaluation was performed using the scoring system and the hip scores were assigned according to the level of pain, the functional status and the range of motion. The patients who refused to return, but who did forward X-rays for review after being contacted were questioned by phone about the functional status of their hip. Radiographic follow-up was performed at six weeks, at three, six and twelve months and yearly thereafter. All the available radiographs were collected and assessed for implant stability, subsidence, osseointegration, osteolysis, stress shielding and evidence of periprosthetic lucency.
Eighty-six hips (78 patients) were available for review at follow-up of greater than 7 years. In 11 of the 86 cases, acetabular failure required revision of the acetabular component, but the femoral stem survived and it was available for long-term evaluation. The radiographs were obtained at 7-year follow-up for another 20 hips, but the patients would not come in for the 7-year clinical evaluation. Therefore, a phone interview was conducted to assess any change in the functional status at a minimum of 7 years.
The mechanical fixation of a tapered, titanium, HA-coated femoral implant was excellent in this study. This femoral design provided reliable osseointegration that was durable at a mean of 7 years follow-up.
骨水泥型髋关节假体的无菌性松动被认为是一个长期存在的问题,尤其是在男性和年轻患者中。人们投入了大量精力开发用于非骨水泥固定的新型假体。我们评估了一种锥形钛羟基磷灰石(HA)涂层股骨髋关节假体的性能和假体周围骨反应,该假体在初次全髋关节置换术后至少 7 年的随访中进行了治疗。
本研究纳入了 78 例患者的 86 髋。其中男性 35 例,女性 43 例;手术时的平均年龄为 59 岁(范围 41 至 81 岁)。我们使用了一种锥形钛(Ti6Al4V)HA 涂层股骨植入物。我们在初次全髋关节置换术后至少 7 年的时间里对患者进行了评估。临床评估采用评分系统进行,髋关节评分根据疼痛程度、功能状态和运动范围进行分配。拒绝返回但在联系后转发 X 射线进行审查的患者通过电话询问其髋关节的功能状态。术后 6 周、3 个月、6 个月和 12 个月以及此后每年进行放射学随访。收集并评估所有可用的 X 射线以评估假体稳定性、下沉、骨整合、骨溶解、应力屏蔽和假体周围透亮区的证据。
86 髋(78 例患者)在随访时间超过 7 年时可进行评估。在 86 例中,有 11 例髋臼失败需要更换髋臼部件,但股骨柄存活且可进行长期评估。另外 20 髋在 7 年随访时获得了 X 射线,但患者不愿进行 7 年的临床评估。因此,进行了电话访谈以评估至少 7 年时功能状态的任何变化。
在这项研究中,锥形钛 HA 涂层股骨植入物的机械固定效果非常好。这种股骨设计提供了可靠的骨整合,在平均 7 年的随访中具有耐用性。