Verma Sumit, Adler Stuart, Berman Adam, Duran Aurillo, Loar Diane
J Interv Card Electrophysiol. 2011 Jan;30(1):37-44. doi: 10.1007/s10840-010-9525-1. Epub 2010 Dec 14.
Transseptal catheterization of the interatrial septum has traditionally been performed under the guidance of fluoroscopy, echocardiography, and hemodynamic pressure monitoring. We hypothesized that the fossa ovalis could be identified on pre-ablation chest computerized tomography (CT) scan utilizing EnSite Verismo and Fusion software thereby permitting its real-time visualization during transseptal puncture.
The purpose of this study is to generate multi-center data demonstrating the feasibility of the identification of the fossa ovalis and visualization of the transseptal needle tip using EnSite Verismo segmentation and EnSite Fusion prior to transseptal puncture.
Patients scheduled to undergo transseptal puncture for ablation were enrolled. Pre-ablation CT scans were performed in all patients. The Verismo software tool was used to import and segmentally manipulate a three-dimensional CT image on the EnSite system workstation. EnSite Fusion registration was used to register the right atrial image and label and confirm the fossa ovalis' location prior to transseptal puncture.
Analysis of the pre-ablation CT scan using EnSite Fusion successfully allowed assessment of the anatomy and location of the fossa ovalis. Additionally, the transseptal needle tip location was located within the CT scan designated fossa ovalis location in 96% of cases.
It is possible to visualize the fossa ovalis on pre-ablation CT and localize the transseptal needle accurately within the margins of the fossa ovalis utilizing EnSite Fusion and Verismo software.
传统上,房间隔穿刺导管插入术是在荧光透视、超声心动图和血流动力学压力监测的引导下进行的。我们推测,利用EnSite Verismo和Fusion软件,可以在消融前胸部计算机断层扫描(CT)上识别卵圆窝,从而在房间隔穿刺过程中实现其实时可视化。
本研究的目的是生成多中心数据,证明在房间隔穿刺前使用EnSite Verismo分割和EnSite Fusion识别卵圆窝并可视化房间隔穿刺针尖端的可行性。
纳入计划接受房间隔穿刺消融的患者。所有患者均进行了消融前CT扫描。使用Verismo软件工具在EnSite系统工作站上导入并分段操作三维CT图像。在房间隔穿刺前,使用EnSite Fusion配准来配准右心房图像并标记和确认卵圆窝的位置。
使用EnSite Fusion对消融前CT扫描进行分析,成功评估了卵圆窝的解剖结构和位置。此外,在96%的病例中,房间隔穿刺针尖端位置位于CT扫描指定的卵圆窝位置内。
利用EnSite Fusion和Verismo软件,可以在消融前CT上可视化卵圆窝,并在卵圆窝边缘内准确确定房间隔穿刺针的位置。