Lee Kye-Sung, Thompson Kevin P, Rolland Jannick P
The Institute of Optics, University of Rochester, 275 Hutchinson Road, Rochester, New York 14627-0186, USA.
Opt Express. 2010 Oct 25;18(22):23378-84. doi: 10.1364/OE.18.023378.
We report an optical design for a low-cost optics, broadband, astigmatism-corrected practical spectrometer. An off-the-shelf cylindrical lens is used to remove astigmatism over the full bandwidth. Results show that better than 0.1 nm spectral resolution and more than 50% throughput were achieved over a bandwidth of 400 nm centered at 800 nm.