Jani S K, Pennington E C
Division of Radiation Oncology, University of Iowa College of Medicine, Iowa City 52242.
Med Phys. 1990 May-Jun;17(3):481-2. doi: 10.1118/1.596488.
Tissue compensators made with a high atomic number material are compact and retain the skin-sparing effect of megavoltage beams. In this note, we have described the use of a commercially available vinyl lead as a material for tissue compensators for use with 4-MV beams on a Clinac-4/80. The dosimetric data obtained included the sheet's lead equivalence, thickness ratios, surface dose and buildup region, and transmission factors. The presence of a compensator did not alter the beam's skin-sparing effects. It was concluded that the vinyl lead sheets allowed an easy and rapid fabrication of a tissue compensator for head and neck portals of up to 15 X 15 cm.
用高原子序数材料制成的组织补偿器结构紧凑,并能保持兆伏级射束的皮肤 sparing 效应。在本报告中,我们描述了使用市售的乙烯基铅作为组织补偿器的材料,用于 Clinac-4/80 上的 4 兆伏射束。获得的剂量学数据包括薄片的铅当量、厚度比、表面剂量和建成区以及透射因子。补偿器的存在并未改变射束的皮肤 sparing 效应。得出的结论是,乙烯基铅薄片能够轻松快速地制造用于头部和颈部最大 15×15 厘米射野的组织补偿器。