Thomas S J, Bruce G
Medical Physics Department, Addenbrooke's Hospital, Cambridge, UK.
Phys Med Biol. 1988 Jun;33(6):703-10. doi: 10.1088/0031-9155/33/6/004.
In radiotherapy treatments with MV beams, the use of tissue compensators affects the dose to the skin. Methods of calculating the relative skin dose (RSD) are described and a formula is derived to predict the contribution of a tissue compensator to the RSD. Measurements of RSD for various field sizes and distances from the compensator are presented.