Suppr超能文献

比较连续患者使用 Xience V 和 Taxus Liberte 支架治疗后的斑块脱垂。

Comparison of plaque prolapse in consecutive patients treated with Xience V and Taxus Liberte stents.

机构信息

Thoraxcenter, Erasmus Medical Center, Ba-583, 's-Gravendijkwal 230, 3015, CE, Rotterdam, The Netherlands.

出版信息

Int J Cardiovasc Imaging. 2012 Jan;28(1):23-31. doi: 10.1007/s10554-010-9768-z. Epub 2010 Dec 21.

Abstract

The purpose of this article is to investigate the prevalence of plaque prolapse (PP) after Xience V and Taxus Liberte stent implantation. During the study period 2006-2007, 200 consecutive patients underwent coronary revascularization for de novo lesions and received an intravascular ultrasound (IVUS) post-stenting evaluation, (n = 124 patients with Taxus Liberte and n = 76 with Xience V) (227 stent segments). Cross-sectional and longitudinal 3D IVUS images were analyzed in a blind fashion, evaluating the prevalence of PP and calculating its depth and angle. The angulation degree of the coronary artery at the lesion site pre-stent implantation was also evaluated by angiography. The prevalence of PP was 23.9% in Xience V versus 38.1% in Taxus Liberte (P = 0.025). The depth and angle of PP were greater in Taxus Liberte stent than Xience V stent (0.4 ± 0.1 mm versus 0.5 ± 0.2 mm, P = 0.004; and 32.0 ± 8.9° versus 44.6 ± 27.6°, P = 0.044, respectively). The angulation degree of the coronary artery at the lesion site was higher in presence of plaque prolapse than in its absence (48.2 ± 29.3° vs. 38.2 ± 28.1°, P = 0.013). By multivariate analysis, stent type was independently associated with incidence of plaque prolapse. Xience V stent has less plaque prolapse than Taxus Liberte stent. Stent design may play a role in the prevalence of plaque prolapse.

摘要

本文旨在探讨 Xience V 和 Taxus Liberte 支架植入术后斑块脱垂(PP)的发生率。在 2006-2007 年的研究期间,200 例接受冠状动脉血运重建术的新发病变患者接受了血管内超声(IVUS)支架置入后评估(n=124 例 Taxus Liberte 支架和 76 例 Xience V 支架)(227 个支架节段)。以盲法分析横断面和纵向 3D-IVUS 图像,评估 PP 的发生率,并计算其深度和角度。支架置入前病变部位冠状动脉的成角程度也通过血管造影进行评估。Xience V 组的 PP 发生率为 23.9%,Taxus Liberte 组为 38.1%(P=0.025)。Taxus Liberte 支架的 PP 深度和角度均大于 Xience V 支架(0.4±0.1mm 比 0.5±0.2mm,P=0.004;32.0±8.9°比 44.6±27.6°,P=0.044)。存在斑块脱垂的病变部位冠状动脉的成角程度高于不存在斑块脱垂的病变部位(48.2±29.3°比 38.2±28.1°,P=0.013)。多变量分析显示,支架类型与斑块脱垂的发生率独立相关。Xience V 支架的斑块脱垂发生率低于 Taxus Liberte 支架。支架设计可能在斑块脱垂的发生率中起作用。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验