Department of Orthopaedics, Oslo University Hospital, Ullevål, Norway.
Acta Orthop. 2011 Feb;82(1):13-9. doi: 10.3109/17453674.2010.548027. Epub 2010 Dec 29.
Plasma-sprayed hydroxyapatite (HA) is a successful coating for fixation of uncemented femoral stems. There may be alternative coatings with advantages in bone remodeling and transport of bone-active substances. We investigated whether an electrochemically deposited hydroxyapatite, Bonemaster (BM), might be a safe alternative in total hip arthroplasty. Our hypothesis was that the new coating would not be inferior to the conventional one.
50 patients (55 hips) were included. The stem was tapered and porous-coated proximally. On top of the porous coating was either HA or BM. Patients were evaluated postoperatively and after 3, 6, 12, and 24 months to measure fixation by radiostereometric analysis (RSA), bone mineral density by dual-energy X-ray absorptiometry (DXA), and conventional radiography. Clinical evaluation was performed with Harris hip score and Oxford hip score, both preoperatively and after 2 years.
After 2 years, the stems had subsided 0.25 (HA) and 0.28 (BM) mm and there were no statistically significant differences between the groups in any direction, regarding both migration and rotation. The BM group retained significantly more bone than the HA group in Gruen zone 1 during the first 2 years. The Harris and Oxford hip scores were similar in both groups.
Electrochemically deposited hydroxyapatite on an uncemented stem does not appear to be inferior to plasma-sprayed HA regarding clinical and radiological results, bone remodeling, and micromotion after 2 years follow-up.
等离子喷涂羟基磷灰石(HA)是固定非骨水泥股骨柄的成功涂层。可能存在具有骨重塑和骨活性物质转运优势的替代涂层。我们研究了电化学沉积羟基磷灰石(Bonemaster,BM)作为全髋关节置换术的一种安全替代物是否可行。我们的假设是,这种新涂层不会逊于传统涂层。
50 例患者(55 髋)纳入研究。柄近端呈锥形,多孔涂层。多孔涂层上覆盖 HA 或 BM。术后、3、6、12 和 24 个月对患者进行评估,通过放射学体层摄影测量分析(RSA)测量固定情况,通过双能 X 射线吸收法(DXA)测量骨密度,并进行常规 X 线摄影。术前和术后 2 年采用 Harris 髋关节评分和 Oxford 髋关节评分进行临床评估。
2 年后,HA 组和 BM 组的柄下沉分别为 0.25mm 和 0.28mm,在任何方向上,两组的迁移和旋转均无统计学差异。在最初 2 年内,BM 组在 Gruen 区域 1 中保留的骨量明显多于 HA 组。两组的 Harris 和 Oxford 髋关节评分相似。
在 2 年随访中,未骨水泥固定柄上电化学沉积的羟基磷灰石在临床和影像学结果、骨重塑和微动方面似乎并不逊于等离子喷涂 HA。