Accurate and comfortable methods are needed to determine resting energy expenditure (REE) in older people who are characterized by a lowered metabolic rate. The portable SenseWear® armband (SWA) body monitor, worn on the right upper arm, can easily be used by this age group in an ambulatory manner.
The purpose of this study was to evaluate the reliability and accuracy of the SWA armband in determining REE in healthy, normal-weight older people.
Participants were 49 older, community-dwelling volunteers aged 60-87 years. Reliability was tested in all subjects. Accuracy was analyzed in 32 subjects by comparing REE estimated by SWA against REE measured by indirect calorimetry as a criterion method. Data were simultaneously and continuously recorded for 20 min in the morning. In the same subjects, REE determined by SWA from night recording was evaluated for accuracy against indirect calorimetry. The agreement between methods was assessed by the Bland-Altman procedure.
(1) REE measured by SWA was reliable when comparing 2 consecutive measurements (typical error: 1.9%). (2) REE calculated from morning recording (1,543 ± 181 kcal/24 h) was higher than REE from night recording (1,564 ± 192 kcal/24 h; p = 0.018). (3) Compared with REE by indirect calorimetry (1,377 ± 228 kcal/24 h) for accuracy, the SWA overestimated mean REE in the older people by 12 and 14%, respectively, for morning and night recording (each p < 0.001).
The SWA provides a reliable estimate of REE in healthy older subjects and has the advantage of easy handling. The 20-min recording time, which was recommended by the manufacturer, can be applied. However, the SWA overestimates REE in this group, possibly due to age-related changes in skin conductance and thermoregulation, both being measured by sensors of the SWA armband. This requires improving the SWA by developing better fitting algorithms for predicting REE in older people.
需要准确且舒适的方法来确定代谢率降低的老年人的静息能量消耗(REE)。可佩戴式 SenseWear®臂带(SWA)身体监测器佩戴在上臂的右侧,该年龄段的人可以轻松地以非卧床的方式使用。
本研究的目的是评估 SWA 臂带在确定健康、正常体重老年人 REE 方面的可靠性和准确性。
参与者为 49 名年龄在 60-87 岁的社区居住的老年志愿者。所有受试者均进行了可靠性测试。32 名受试者通过比较 SWA 估计的 REE 与间接测热法作为标准方法测量的 REE 来分析准确性。数据在早上同时连续记录 20 分钟。在同一组受试者中,评估 SWA 从夜间记录中确定的 REE 对间接测热法的准确性。使用 Bland-Altman 程序评估方法之间的一致性。
(1)当比较 2 次连续测量时,SWA 测量的 REE 是可靠的(典型误差:1.9%)。(2)从早晨记录计算的 REE(1,543 ± 181 kcal/24 h)高于夜间记录(1,564 ± 192 kcal/24 h;p = 0.018)。(3)与间接测热法的 REE(1,377 ± 228 kcal/24 h)相比,SWA 分别高估了老年人早晨和夜间记录时 REE 的平均值 12%和 14%(p < 0.001)。
SWA 可提供健康老年受试者 REE 的可靠估计值,具有易于操作的优点。制造商推荐的 20 分钟记录时间可以应用。然而,SWA 高估了该组的 REE,这可能是由于年龄相关的皮肤电导和体温调节变化,这两者都是由 SWA 臂带的传感器测量的。这需要通过开发更好的适用于预测老年人 REE 的拟合算法来改进 SWA。