Evaluation of a Portable Armband Device to Assess Resting Energy Expenditure in Patients With Anorexia Nervosa.

作者信息

Elbelt Ulf, Haas Verena, Hofmann Tobias, Stengel Andreas, Berger Heike, Jeran Stephanie, Klapp Burghard F

机构信息

Department of Endocrinology, Diabetes and Nutrition, Charité-Universitätsmedizin Berlin, Berlin, Germany Charité Center for Internal Medicine and Dermatology, Division of General Internal and Psychosomatic Medicine, Charité-Universitätsmedizin Berlin, Berlin, Germany

Charité Center for Internal Medicine and Dermatology, Division of General Internal and Psychosomatic Medicine, Charité-Universitätsmedizin Berlin, Berlin, Germany Department of Child and Adolescent Psychiatry, Charité-Universitätsmedizin Berlin, Berlin, Germany.

出版信息

Nutr Clin Pract. 2016 Jun;31(3):362-7. doi: 10.1177/0884533615618900. Epub 2015 Dec 1.

Abstract

BACKGROUND

In women with anorexia nervosa (AN), resting energy expenditure (REE) is decreased due to reduced energy intake and severe underweight. The assessment of REE allows estimating individual metabolic downregulation and better understanding body weight regulatory mechanisms in severely underweight patients with AN. However, REE predictive equations are known to have considerable shortcomings in patients with AN. Our aim was to evaluate a portable armband device (SenseWear armband [SWA]; BodyMedia, Inc, Pittsburgh, PA) for the assessment of REE against the measurement with indirect calorimetry (IC) as the reference method.

METHODS

We assessed REE simultaneously by IC and SWA in 50 women with AN at the start of inpatient therapy and calculated REE using 2 predictive equations.

RESULTS

Reliable data for IC measurement were obtained for 34 patients (age: 27.0 ± 8.0 years; body mass index: 14.4 ± 2.0 kg/m²). REE assessed with SWA was overestimated by 23% ± 27% compared with REE measured by IC (1166 ± 174 vs 979 ± 198 kcal/d, P < .001). REE estimation with SWA gave an accurate prediction within 10% deviation of REE measured with IC in 35% of the patients. In contrast, REE calculated with 2 predictive equations underestimated REE measured with IC by -26% ± 17% and -5% ± 20%, respectively.

CONCLUSIONS

A mean difference of 187 kcal/d between both techniques for the assessment of REE may be of methodological relevance. Therefore, SWA and IC are not interchangeable methods for the assessment of REE in underweight females with AN.

摘要

背景

在神经性厌食症(AN)女性患者中,由于能量摄入减少和严重体重过轻,静息能量消耗(REE)会降低。评估REE有助于估计个体代谢下调情况,并更好地理解严重体重过轻的AN患者的体重调节机制。然而,已知REE预测方程在AN患者中存在相当大的缺陷。我们的目的是评估一种便携式臂带设备(SenseWear臂带[SWA];BodyMedia公司,宾夕法尼亚州匹兹堡)用于评估REE,并以间接测热法(IC)测量作为参考方法。

方法

我们在50名AN女性患者住院治疗开始时,同时通过IC和SWA评估REE,并使用2个预测方程计算REE。

结果

34例患者(年龄:27.0±8.0岁;体重指数:14.4±2.0kg/m²)获得了可靠的IC测量数据。与IC测量的REE相比,SWA评估的REE高估了23%±27%(分别为1166±174 vs 979±198kcal/d,P<.001)。在35%的患者中,SWA对REE的估计在与IC测量的REE偏差10%以内给出了准确预测。相比之下,用2个预测方程计算的REE分别比IC测量的REE低估了-26%±17%和-5%±20%。

结论

两种评估REE技术之间平均相差187kcal/d可能具有方法学相关性。因此,SWA和IC不是评估体重过轻的AN女性患者REE的可互换方法。